This can be confirmed with cursors:
The hysteresis value is used to eliminate noise. A non-monotonicity is
detected only when its amplitude is greater than the hysteresis. Therefore,
when setting a hysteresis level, set a value that is greater than the
amplitude of the noise.

Runt Mode

This search mode looks for pulses that fail to transit a given region. You
can search for positive-going or negative-going runts, or both. An
adjustable hysteresis band is provided to eliminate noise.
In the case of negative-going runt pulses, the value displayed in the table is
the difference (delta) of the high level of the signal and the runt amplitude,
i.e., where the runt bottoms out.