Vertical Setup; Coupling; Deskew; Probe Attenuation - LeCroy WaveSurfer MXs-B Getting Started Manual

Mxs-b series oscilloscopes
Hide thumbs Also See for WaveSurfer MXs-B:
Table of Contents

Advertisement

G
S
ETTING
TARTED
At this point, the following dialog is shown at the bottom of the display, and the
grid area shortens until this dialog is closed (by touching the Close button on the
upper right hand corner of this dialog).
V
S
ERTICAL
ETUP

Coupling

Coupling choices are as follows:
DC 50 Ω
GROUND
DC 1 MΩ
AC 1 MΩ
Select coupling by touching inside the Coupling field and choosing a coupling
mode from the pop-up menu.
Note: The coupling choices for a channel change if a ProBus probe is connected to
a channel.

Deskew

Deskew allows you to compensate for different lengths of cables, probes, or
anything else that might cause timing mismatches between signals. Connect all
probes to the desired channels, then probe a common signal with each probe and
adjust for timing differences using Deskew.

Probe Attenuation

If you use a LeCroy ProBus compatible active probe, or a probe compatible with
Probe Ring, the attenuation is automatically set by the oscilloscope. If it is not
automatically set, select a value here by touching inside the Probe Atten field
selecting a value from the pop-up menu.
33
M
ANUAL
WSMXs-B-GSM-E Rev A

Advertisement

Table of Contents
loading

Table of Contents