Getting Started M anual WavePro 7Zi and Zi-A Oscilloscopes...
Page 2
WavePro 7 Zi and Zi-A Oscilloscopes November, 2010...
Page 3
LeCroy shall not be responsible for any defect, damage, or failure caused by any of the following: a) attempted repairs or installations by personnel other than LeCroy representatives or b) improper connection to incompatible equipment, or c) for any damage or malfunction caused by the use of non-LeCroy supplies.
Getting Started Manual TABLE OF CONTENTS Welcome ..............12 Support ..............12 Thank You ..............13 Safety Requirements ..........13 Safety Symbols ............13 Operating Environment ..........15 Safety Certification ........... 16 Cooling ..............17 AC Power Source............17 Power Consumption ..........18 Power and Ground Connections .......
Page 5
WavePro 7 Zi and Zi-A Oscilloscopes Vertical Front Panel Controls ........34 Cursors Front Panel Controls ........35 WaveScan Front Panel Controls ....... 36 General Controls Front Panel Controls ....36 Zoom and Math Front Panel Controls ...... 37 Connecting ..............38 Probe Interfaces ............
Page 7
WavePro 7 Zi and Zi-A Oscilloscopes Inverting Waveforms ..........73 Finding Scale............. 73 Using Find Scale............73 Variable Gain ............74 Enabling Variable Gain ..........74 Channel Deskew ............74 Channel Deskew Setup ..........74 Response Optimization Modes ........ 74 Trigger ...............
Page 8
Getting Started Manual Viewing Waveforms ........... 92 Display ............... 92 Display Setup ............92 Moving Traces from Grid to Grid ......94 XY Display ..............95 Display Grids ............. 96 Zooming Waveforms ..........96 Zooming a Single Channel ......... 97 Touch-and-Drag Zooming .........
Page 9
WavePro 7 Zi and Zi-A Oscilloscopes Analysis..............124 WaveScan ............... 124 Parameter Analysis ..........126 Pass/Fail Testing ............. 128 Customization ............135 Customization Overview ........135 Documenting Your Work Using LabNotebook .... 137 Introduction to LabNotebook ........ 137 Creating a Notebook Entry ........137 Saving and Recalling Setups ........
Page 10
Getting Started Manual Adding Printers and Drivers ........155 Printing a Screen Image .......... 156 Printing..............157 Clipboard ..............157 File ................158 E-mail ..............160 Auxiliary Output Signals .......... 161 Auxiliary Output Setup..........162 Date and Time ............163 Setting the Date and Time Manually ......
Page 11
WavePro 7 Zi and Zi-ASpecifications ...... 174 CERTIFICATIONS ............. 175 CE Declaration of Conformity......... 175 EMC Directive ............175 Low-Voltage Directive ..........176 UL and cUL.............. 176 Contact LeCroy for Support ........177 ® Windows License Agreement ....... 190 Index ................ 191 WP7Zi-GSM-E RevB...
When your WavePro 7 Zi and Zi-A Series Oscilloscope is delivered, verify that all items on the packing list or invoice copy have been shipped to you. Contact your nearest LeCroy customer service center or national distributor if anything is missing or damaged. If you do not contact us immediately, we cannot be responsible for replacement.
WavePro 7 Zi and Zi-A Oscilloscopes Thank You We truly hope these materials provide increased comprehension when using LeCroy's fine products. Sincerely, David C. Graef LeCroy Corporation Vice President and Chief Technology Officer Safety Requirements This section contains information and warnings that must be observed to keep the instrument operating in a correct and safe condition.
Page 14
Getting Started Manual This symbol is used to denote a safety ground connection. On (Supply). This is the AC mains connect/disconnect switch at the back of the instrument. Off (Supply). This is the AC mains connect/disconnect switch at the back of the instrument.
WavePro 7 Zi and Zi-A Oscilloscopes WARNING The WARNING sign indicates a potential hazard. It calls attention to a procedure, practice or condition which, if not followed, could possibly cause bodily injury or death. If a WARNING is indicated, do not proceed until its conditions are fully understood and met.
Getting Started Manual CAUTION Protect the oscilloscope’s display touch screen from excessive impacts with foreign objects. CAUTION Do not exceed the maximum specified front panel terminal (CH1, CH2, CH3, CH4, AUX IN, and AUX OUT) voltage levels. Refer to Specifications for more details.
WavePro 7 Zi and Zi-A Oscilloscopes Protection Class 1 refers to grounded equipment, in which protection against electric shock is achieved by Basic Insulation and by means of a connection to the protective ground conductor in the building wiring. Cooling The instrument relies on forced air cooling with internal fans and ventilation openings.
Getting Started Manual Power Consumption </= 800 watts (800 VA) for all WavePro 7 Zi and Zi-A series models (depending on accessories installed - probes, PC port plug-ins, etc.). Power consumption in Standby Mode: 15 Watts. Power and Ground Connections The instrument is provided with a 10A/250V 18AWGrated grounded cord set containing a molded three-terminal polarized plug and a standard IEC320 (Type C13) connector for making line voltage and safety ground...
Page 19
WavePro 7 Zi and Zi-A Oscilloscopes Standby (Power) Switch and Oscilloscope Operational States The front Standby (Power) switch controls the operational state of the oscilloscope. This toggle switch is activated by momentarily pressing and releasing it. The color of the LED below the switch indicates the status of the oscilloscope as follows: ...
Getting Started Manual The oscilloscope can always be placed in the Standby state (LED Off) – Power Off (except for some housekeeping circuits) by pressing and holding in the Standby toggle switch for approximately 5 seconds. CAUTION Hibernate mode is not supported by the Power button. Therefore, Windows®...
Page 21
WavePro 7 Zi and Zi-A Oscilloscopes Proper use of the instrument depends on careful reading of all instructions and labels. WARNING Any use of the oscilloscope in a manner not specified by the manufacturer may impair the instrument’s safety protection. WP7Zi-GSM-E RevB...
Getting Started Manual Hardware Front Panel Numbered labels on this image correspond with descriptions on the following table. Number and Description 1. Power 5. Volume Control 9. USB Ports Button and Mute Button 10. Detachable Front 2. Channel Row 6. Ground Panel-Remote LED Indicator Connector...
WavePro 7 Zi and Zi-A Oscilloscopes Detaching and Attaching the Front Panel Detach the front panel from the oscilloscope by sliding the detachment lever to the left and pulling at the right. Attach the front panel by inserting the lower part first, sliding the detachment lever to the left, and then pushing the top in place.
Page 24
Note: While a standard front panel comes with your Zi oscilloscope, LeCroy offers additional standard front panels or a 4 channel version (as follows) to better suit the way you work. Learn more and Contact LeCroy for Support (on page 177).
WavePro 7 Zi and Zi-A Oscilloscopes I/O Panel The I/O Panel is located on the right side (facing the front of the instrument) as follows: Numbered labels on this image of the I/O Panel correspond with descriptions on the following table. WP7Zi-GSM-E RevB...
Page 26
Getting Started Manual The exact location does vary on WavePro 7 Zi, Zi-A and WaveMaster 8 Zi, Zi-A instruments. Number and Description 1. LBUS (LeCroy 5. External VGA 10. PCI Expansion Slots Bus) Monitor for DVI (for 2. Mouse 6. Ethernet Port standard display 3.
WavePro 7 Zi and Zi-A Oscilloscopes Back Panel Numbered labels on this image correspond with descriptions on the following table. Number and Description 1. 10 MHz 2. 10 MHz Reference 4. DVD-CD + R Reference Clock Output Drive Clock Input 3.
Getting Started Manual External Display Setting up the External Display involves a connection and a few touch screen selections. It's a hot swap connection, so there's no need to restart the instrument once you've connected. 1. Plug your external display (ZI-EXTDISP-15 option) power cord into an AC socket and the DVI plug into the PCI slot.
Page 29
WavePro 7 Zi and Zi-A Oscilloscopes The following picture shows a WavePro 7 Zi with the LeCroy external display attached (optional ZI-EXTDISP-15). WP7Zi-GSM-E RevB...
Getting Started Manual Basic Controls Hardware and Software Controls The following Basic Control topics cover the general usage of the hardware buttons located on the oscilloscope's front panel and the screen control interface elements of the software. Front Panel Controls Note: Many front panel controls directly correspond with screen layout controls.
Miscellaneous Setup Controls Help - Press to open the LeCroy Online Assistant where you can click to open the oscilloscope online help table of contents, index, or search for a topic using a keyword. If the second monitor is installed, the online help opens on the second monitor.
Getting Started Manual Press the A front panel button twice to perform the last ETUP selection from the Auto Setup flyout menu (the default is to perform a full auto setup). If Auto Setup is run when no channels are turned on, all channels are affected.
WavePro 7 Zi and Zi-A Oscilloscopes Auto - Press to turn on Auto Trigger mode, which triggers the oscilloscope after a time-out, even if the trigger conditions are not met. Normal - Press to turn on Normal Trigger mode, which triggers the oscilloscope each time a signal is present that meets the conditions set for the type of trigger selected.
Getting Started Manual Vertical Front Panel Controls Note: Vertical front panel controls correspond with screen menu selection: Vertical → Channel Setup. Channels - The channel buttons control both channel ON/OFF and which channel is active for the Vertical Offset and Volts/Div knobs controls.
WavePro 7 Zi and Zi-A Oscilloscopes Cursors Front Panel Controls Note: Cursors front panel controls correspond with screen menu selection: Cursors → Cursors Setup. Absolute Cursor - Press to set the cursor position to the default 25% position, either horizontal or vertical (depending on which cursor type you are using).
Getting Started Manual WaveScan Front Panel Controls Note: WaveScan front panel controls correspond with screen menu selection: Analysis → WaveScan. Scroll - If WaveScan is ON, turn to scroll through the table of WaveScan-filtered events. If WaveScan is OFF, turn to adjust the selected data entry field.
WavePro 7 Zi and Zi-A Oscilloscopes Press the Save Screen to Clipboard button on the flyout menu to save the current screen image to the clipboard. Press the Send Screen to Email button to send the current screen image using the email settings on the Utilities → Utilities Setup → Hardcopy dialog.
Furthermore, it provides automatic matching of probe to oscilloscope response - ensuring the most pristine signal fidelity. The LeCroy WavePro 7 Zi and Zi-A series oscilloscopes with bandwidths of 4 to 6 GHz include both ProBus and ProLink probe interfaces.
Page 39
As shown in the following figure, each channel has a high bandwidth ProLink connector and a 50 Ω/1 MΩ ProBus connector. Figure 4-1. A LeCroy Zi series oscilloscope showing the four sets of ProBus/ProLink probe interfaces. Zi oscilloscopes are compatible with all LeCroy probes.
Page 40
Getting Started Manual ProBus also includes sense rings for detecting passive probes. This is a feature unique to LeCroy in the 1.5 - 6 GHz bandwidth range. The following figure shows a typical channel setup. The input selection is on the left hand side of the dialog box. In this case the A input, the ProLink interface is selected.
3.5 GHz range and a dedicated high-bandwidth interface in the range of 4 through 20 GHz. ProLink Interface Adapters For some instruments, LeCroy's ProLink Adapters (LPA) provides the ability to connect your signal in the following ways: Using a BMA connector.
Page 42
Getting Started Manual An LPA-2.92 mm to ProLink Probe Adapter is also available. As its name implies, this adapter converts a WL-2.92MM platform/cable assembly to the LeCroy ProLink interface. Figure 4-4. BMA-to-SMA, BMA-to-BNC, BMA-to-2.92 mm, and LPA-2.92 mm to ProLink Adapters.
WavePro 7 Zi and Zi-A Oscilloscopes Connecting the Adapters The mating end of the ProLink adapter has four fastening clips (shown right). When installing an adapter on the instrument's connector panel, align the male 6-pin connector with the female connector and push the adapter straight in.
Getting Started Manual DC Level - This is a reference level. Trigger Enabled - This function can be used as a gating function to trigger another instrument when the oscilloscope is ready. Pass/Fail - Allows you to set a pulse duration from 1 µs to 131 ms; generates a pulse when pass/fail testing is active and conditions are met.
WavePro 7 Zi and Zi-A Oscilloscopes Probes LeCroy offers a variety of passive and active probes for use with your X- Stream oscilloscope. Visit www.lecroy.com for specifications and ordering information. Current Probes - Current Probes measure the current passing through a wire;...
Getting Started Manual Passive Probe Compensation Passive probes must be compensated to flatten overshoot. This is accomplished by means of a trimmer at the connector end of the probe. 1. Attach the connector end of your passive probe to any channel. 2.
WavePro 7 Zi and Zi-A Oscilloscopes Display Dashboard Screen Layout, Groupings, and Controls The instrument's screen is divided into five main sections: Menu Bar Signal Display Grid Descriptor Labels Dialog(s) The Message Bar Note: Many front panel controls directly correspond with screen layout controls.
Getting Started Manual Specific Menu Bar functions are referenced using arrow-separated path descriptions. For example, the Save Setup function is referenced as File → Save Setup. PLEASE NOTE THE FOLLOWING: For common oscilloscope operations, you don’t need to use the top menu bar (since you can access most dialogs from the Front Panel or from the Descriptor Labels).
WavePro 7 Zi and Zi-A Oscilloscopes An Undo button is often available to the right of the Quick Access Toolbar on the Menu Bar. The Signal Display Grid The grid area is divided into 8 vertical divisions and 10 horizontal divisions just like any other oscilloscope.
Page 50
Getting Started Manual All trigger delay values (including post-trigger delay, shown here) are displayed in the Timebase Descriptor Label. Zero delay is the horizontal center of the oscilloscope display. The default setting (Time) is for delay readout (in seconds) and to move proportionately when the timebase knob is turned.
WavePro 7 Zi and Zi-A Oscilloscopes Signal Display Grid Pop-Up Menu On the Signal Display Grid, the Pop-up menu provides assistance while using the oscilloscope. Clicking on a waveform opens a pop-up menu. From this pop-up menu, you can perform the following functions: ...
Page 52
Getting Started Manual These indicators have a long and short form, respectively. Figure 4-7. The long and short forms of trace descriptor indicators. Besides channel traces, math and parameter measurement labels are also displayed. Labels are displayed only for traces that are turned on. Vertical and horizontal trace descriptor (labels) are displayed below the grid.
WavePro 7 Zi and Zi-A Oscilloscopes Display Pop-Up Menu You can click on a trace descriptor label to open a pop- up menu. From this pop-up menu, you can perform the following functions: Open the Setup dialog for the trace. ...
Page 54
Getting Started Manual From this pop-up you can edit existing annotations, change the label placement on the waveform, add labels, remove labels, and toggle the visibility. The following bullets show you how. If you are modifying an existing label, under Labels touch the label you want to change.
WavePro 7 Zi and Zi-A Oscilloscopes O.K. on the keyboard when you are done. The edited text automatically appears in the label on the waveform. Precisely place the label by touching inside the Horizontal Pos. field and provide a horizontal value, using the pop-up numeric keypad. ...
Getting Started Manual Tool Bar Buttons Several dialogs contain common functions accessible from a row of buttons that can save you a step or two when using a particular function (instead of accessing the respective location from the menu bar). For example, at the bottom of the Channel Setup dialog, these buttons (sometimes labeled as Actions) perform the following functions: ...
WavePro 7 Zi and Zi-A Oscilloscopes Probe Cal - Cable Deskew - Opens the Probes Cal. dialog where various Gain, Offset, Skew, Source, and Advanced controls are available for probe signal calibration. Another example is seen in the buttons appearing at the bottom of the Measure Px dialogs.
Getting Started Manual You can also quickly create traces (and turn on the trace label) for math functions and memory traces, without leaving the Vertical Adjust dialog, by touching the icons at the bottom of the Vertical Adjust dialog. Whenever you turn on a channel, math, or memory trace using either the menu bar or trace descriptor label, the dialog at the bottom of the screen automatically switches to the vertical setup or math setup dialog for that...
WavePro 7 Zi and Zi-A Oscilloscopes 3. Touch inside the Delay data entry field and enter a value, or use the slider bar at the bottom of the window. Click the keypad button (as follows) on the slider bar to enter a value using the pop-up numeric keypad.
Getting Started Manual Refer to Acquisition Modes in the specifications on www.lecroy.com maximum sample rates. OMBINING HANNELS ROCEDURE Set up channel combinations on the interface as follows: 1. Touch Timebase → Horizontal Setup on the menu bar and the Horizontal Setup dialog is shown.
Page 61
WavePro 7 Zi and Zi-A Oscilloscopes The acquisition is typically stopped a defined number of samples after this event occurs: a number determined by the selected trigger delay and measured by the timebase. The waveform's horizontal position (and waveform display in general) is determined using the trigger event as the definition of time zero.
Page 62
Getting Started Manual You can also make time measurements between events on selected segments using the full precision of the acquisition timebase. Sequence mode offers a number of unique capabilities: You can acquire up to four channels simultaneously. You can minimize dead time between trigger events for consecutive segments.
WavePro 7 Zi and Zi-A Oscilloscopes Sequence Display Modes The instrument gives you a choice of five ways to display your segments: Adjacent Waterfall (cascaded) Mosaic (tiled) Overlay WP7Zi-GSM-E RevB...
Getting Started Manual Perspective Note: some display modes have limitations on the number of segments that can be shown at one time. Sequence Mode Setup When setting up Sequence Mode, you define the number of fixed-size segments acquired in single-shot mode (see the instrument specifications for the limits).
Page 65
WavePro 7 Zi and Zi-A Oscilloscopes 1. Touch Timebase → Horizontal Setup on the menu bar. 2. Click the Sequence tab. 3. Under Acquisition Settings, touch inside the Num Segments data entry field and enter the number of segments you want to display, using the slider bar at the bottom of the window.
Getting Started Manual 6. Touch the S trigger front panel button. INGLE Note: Once a single acquisition has started, you can interrupt the acquisition at any time by pressing the S front panel button a INGLE second time or by pressing the S front panel button.
WavePro 7 Zi and Zi-A Oscilloscopes 2. Turn the Z front panel position knob to scroll through the segments. 3. To vary the degree of zoom, touch the newly created Zx trace label. The setup dialog for the zoom (Z1 to Z4) opens. It shows the current horizontal and vertical zoom factors.
Getting Started Manual 5. Touch inside the First Selected data entry field and use the slider bar at the bottom of the window to select the first segment you want to display. Click the keypad button (as follows) on the slider bar to enter a value using the pop-up numeric keypad.
WavePro 7 Zi and Zi-A Oscilloscopes 4. Touch inside the Select Segment field and enter a segment number, using the pop-up keypad or touch the arrow buttons to scroll through segment times. RIS Sampling Mode - For Higher Sampling Rates RIS (Random Interleaved Sampling) is an acquisition technique that allows effective sampling rates higher than the maximum single-shot sampling rate.
Getting Started Manual Roll Mode Roll mode displays, in real time, incoming points in single-shot acquisitions that have a sufficiently low data rate. The oscilloscope rolls the incoming data continuously across the screen until a trigger event is detected and the acquisition is complete.
WavePro 7 Zi and Zi-A Oscilloscopes The set voltage is shown on the trace descriptor label and in the Volts/Div field in the dialog. Adjusting the Waveform's Position Turn the V adjust front panel knob directly above the channel ERTICAL FFSET button whose waveform you want to move vertically.
Probe Attenuation Setting up Probe Attenuation LeCroy's ProBus system automatically senses probes and sets their attenuation for you. If you want to set the attenuation manually, 1. In the menu bar, touch Vertical, and then select a channel from the drop-down menu.
WavePro 7 Zi and Zi-A Oscilloscopes Anti-aliasing filters are applied to limit bandwidth at lower sample rates, regardless of bandwidth limit selection. Linear and (SinX)/X Interpolation Linear interpolation, which inserts a straight line between sample points, is best used to reconstruct straight-edged signals such as square waves. (Sinx)/x interpolation, on the other hand, is suitable for reconstructing curved or irregular wave shapes, especially when the sample rate is 3 to 5 times the system bandwidth.
Getting Started Manual Variable Gain Variable Gain lets you change the granularity with which the gain is incremented. For example, when Variable Gain is disabled, the gain increases or decreases in preset increments each time you touch the Up/Down buttons. However, when Variable Gain is enabled, you can increase or decrease the gain in increments as small as 0.1 mV/div, depending on the waveform scale.
Page 75
Then, by providing the flattest possible frequency response with zero group delay, the fastest possible rise time can be measured. LeCroy provides all three of these choices in the Response Optimization Mode selection. Pulse Response - A delay compensation that minimizes preshoot.
Getting Started Manual Flatness - A delay compensation similar to eye diagram, this selection provides a flat frequency response and group delay. It provides the fastest rise time. Trigger Trigger Overview Triggering is the way an oscilloscope selects an exact moment in time on a signal to then be shown on the screen.
In addition, there are more sophisticated triggers for setting basic or complex conditions. These include Smart Triggers, Serial Trigger, Qualified Triggers, and Cascade Triggers. Learn more and Contact LeCroy for Support (on page 177). Edge A simple trigger, Edge trigger is activated by basic waveform features or conditions such as positive or negative slope, and holdoff.
Getting Started Manual If the oscilloscope is in Normal trigger mode, it automatically resets after the B event. Arm trigger events can be set for Edge, Pattern, State, and PatState. State is briefly explained as follows: TATE The State trigger is a level-qualified trigger which requires that the qualifying signal remain above or below a specified voltage level for a trigger to occur.
Page 79
WavePro 7 Zi and Zi-A Oscilloscopes LITCH Glitch trigger is a simpler form of Width trigger. Use Glitch trigger when you want to define a fixed pulse-width time or time range only. Glitch trigger makes no provision for voltage levels or ranges. ROPOUT Used primarily in single-shot applications, and usually with a pre-trigger delay, Dropout trigger can detect lost signals.
Getting Started Manual Pattern (Logic) Logic trigger enables triggering on a logical combination (pattern) of five inputs: CH1, CH2, CH3, CH4, and AUX. You have a choice of four Boolean operators (AND, NAND, OR, NOR), and you can stipulate the high or low voltage logic level for each input independently.
WavePro 7 Zi and Zi-A Oscilloscopes Serial Trigger Protocol-enabled serial trigger capabilities are available as options on several oscilloscope models. Clock and Data Recovery (CDR) 80-bit Serial Trigger without decode capability is standard on SDA model oscilloscopes and optional on WaveMaster and DDA model oscilloscopes. Trigger Settings Horizontal and Vertical adjustments are typically made for all trigger types using either the Delay or Level knobs on the front panel of the instrument...
Getting Started Manual Vertical: Turn the L knob in the TRIGGER control group to adjust EVEL the trigger's vertical threshold. Turn this knob to adjust the level of the trigger source or the highlighted trace. Level defines the source voltage at which the trigger will generate an event: a change in the input signal that satisfies the trigger conditions.
Page 83
WavePro 7 Zi and Zi-A Oscilloscopes Touch inside the trigger Source field and select a source on which to trigger. If available, touch inside the Coupling field and select a coupling mode. Coupling refers to the type of signal coupling at the input of the trigger circuit.
Page 84
Getting Started Manual Touch inside the trigger Slope field and choose the direction of the trigger voltage transition used for generating a particular trigger event. The selection is then shown to the right of the dialog as follows for a Positive Slope selection on an Edge Trigger.
Page 85
WavePro 7 Zi and Zi-A Oscilloscopes The selected trigger level is associated with the chosen trigger source. Note: The trigger level is specified in volts and normally remains unchanged when the vertical gain or offset is modified. The Amplitude and Range of the trigger level are limited as follows: ...
Page 86
Getting Started Manual IDTH ONDITION IS Width triggers have additional Width Condition settings as follows. Your Width Condition can be specified as Less Than, Greater Than, In Range and Out Of Range. In Range and Out Of Range conditions can be further set with Upper and Lower Values (for Limits).
Page 87
WavePro 7 Zi and Zi-A Oscilloscopes Now, click the corresponding field and provide a value using the slider bar at the bottom of the window. Click the keypad button (as follows) on the slider bar to enter a value using the pop-up numeric keypad. Use holdoff to obtain a stable trigger for repetitive, composite waveforms.
Page 88
Getting Started Manual Hold Off by Events Select a positive or negative slope and a number of events. An event is the number of times the trigger condition is met after the last trigger. A trigger is generated when the condition is met after this number, counted from the last trigger.
WavePro 7 Zi and Zi-A Oscilloscopes TriggerScan TriggerScan is a debugging tool that helps you quickly find rare waveform glitches and anomalies. With TriggerScan, you can build a list of trigger setups to look for rare events and automatically sequence through each one.
Getting Started Manual Training TriggerScan The TriggerScan Trainer inspects a currently acquired waveform and automatically builds a list of common trigger setups used to find rare events. PLEASE NOTE THE FOLLOWING: You must acquire and display at least 3 cycles of a signal before running the Trainer.
WavePro 7 Zi and Zi-A Oscilloscopes Starting TriggerScan After you have run the Trainer, the Trigger List displays a list of smart trigger setups. You can add or remove trigger setups. You can also update the selected smart trigger setup. Once you have made any changes to the Trigger List, you are ready to start scanning.
Getting Started Manual If you want TriggerScan to stop when the oscilloscope triggers next, check the Stop On Trigger checkbox. You can use this to isolate specific trigger setups. Saving TriggerScan Setups You should save TriggerScan setups once you have made any modifications to the Trigger List.
Page 93
WavePro 7 Zi and Zi-A Oscilloscopes 3. Touch inside the Grid Intensity data entry field (as follows). Provide a value from 0 to 100 using the slider bar at the bottom of the window. Click the keypad button (as follows) on the slider bar to enter a value.
Getting Started Manual 5. Touch inside the trace Intensity data entry field if you want to change the value. Enter a value from 0 to 100 using the slider bar at the bottom of the window. Click the keypad button (as follows) on the slider bar to enter a value.
WavePro 7 Zi and Zi-A Oscilloscopes 2. Touch the Next Grid button. Note: If you have more than one waveform displayed on only one grid, a second grid automatically opens when you select Next Grid. XY Display Use XY displays to measure the phase shift between otherwise identical signals.
Getting Started Manual Display Grids Some instruments and software packages offer the option to display side- by-side, instead of vertically stacked grids. These side-by-side grids are in the Display dialog. Included in the choices are Auto, Single, Dual, Quad, Octal, XY, XY Single, XY Dual, Tandem, Triple, Quattro, and Hex displays. These additional grid displays offer more convenient views of multiple waveforms.
WavePro 7 Zi and Zi-A Oscilloscopes You can also use the front panel Q button to create multiple UICK zooms, one for each displayed input channel. For more information, see the To Zoom Multiple Waveforms Quickly (on page 100) topic. REVIEWING OOMED AVEFORMS...
Page 98
Getting Started Manual 2. Touch the Zoom at the bottom of the Cx Vertical Adjust dialog. A zoom trace (one of Z1 to Z4) is created for the selected channel. 3. Vary the degree of zoom by touching the newly created Zx trace label.
WavePro 7 Zi and Zi-A Oscilloscopes Horizontal Ratio - Press to toggle between fixed and variable horizontal zoom ratio adjustment. Turn to change the horizontal zoom ratio of the selected math trace. Quick Zoom - Press to automatically display magnified views of up to four signal inputs on multiple grids.
Getting Started Manual Quickly Zooming Multiple Waveforms Press the front panel Q button. UICK Turning Off Zoom 1. Touch the math function trace label (or Zx trace label) for the zoom you want to turn off. 2. Touch the Trace On checkbox to remove the check mark and disable the zoom trace.
Page 101
WavePro 7 Zi and Zi-A Oscilloscopes Persistence Setup for All Input Channels at Once 1. Touch the All Locked button. Touch one of the Mode buttons. 3. Touch the Show last trace checkbox if you want the last trace displayed. 4.
Getting Started Manual To turn off persistence for an individual channel, touch the left- most persistence mode button next to the channel. For each input channel, touch its tab to set up persistence for that channel. Touch one of the mode buttons. 4.
WavePro 7 Zi and Zi-A Oscilloscopes WaveStream mode operates at up to 80 GS/s with an update rate of up to several thousand waveforms/second for better capture of higher frequency abnormal events. Adjusting Trace Intensity The I knob adjusts the brightness of your trace. NTENSITY The intensity value is shown in the bottom-right corner of the screen.
Getting Started Manual Click the keypad button (as follows) on the slider bar to enter a value using the pop-up numeric keypad. Analyzing Waveforms Measure Measuring with Cursors Cursors are important tools that aid you in measuring signal values. Cursors are markers - lines, cross-hairs, or arrows - that you can move around the grid or the waveform itself.
WavePro 7 Zi and Zi-A Oscilloscopes Delta Displays the difference in Y values. The value can be read in the descriptor label for the trace. Abs+Delta Displays absolute and delta cursors together. Slope Displays the slope between cursors. If there are non–time-domain waveforms displayed, there will also be a menu offering choices of x-axis units: s or Hz, for example.
Page 106
Getting Started Manual Press the C button on the front panel repeatedly to scroll URSOR through each cursor type. Turn the C knobs to adjust the URSORS position of the cursor. Note: The bottom cursor knob adjusts relative cursors. It does not work with absolute cursor types.
WavePro 7 Zi and Zi-A Oscilloscopes Click the keypad button (as follows) on the slider bar to enter a value using the pop-up numeric keypad. Use the C knobs on the front panel to place the cursors. URSORS 6. If you want both cursors to move in unison as you adjust the position, touch the Track check box to enable tracking.
Getting Started Manual 2. Touch the On checkbox for each parameter you want to enable. 3. Touch the Show Table checkbox to display the parameters below the grid. UICK CCESS TO ARAMETER ETUP IALOGS You can quickly gain access to a parameter setup dialog by touching the parameter list box below the grid.
Page 109
Source1 and Source2 fields. Advanced Web Edit Selecting this type uses LeCroy's Processing Web for measurement setup. This feature, available with the XWEB option, allows you to chain practically unlimited math functions together for operation on your waveform measurements.
Getting Started Manual If you selected Math On Parameters, touch inside the Math Operator data entry field and select a math function from the pop- up menu. 5. In the mini-dialog to the right of the main setup dialog, touch the Gate tab to narrow the span of measurement - as covered in the Measure Gate topic (available in the online help files).
WavePro 7 Zi and Zi-A Oscilloscopes TANDARD ORIZONTAL ARAMETERS These are the default Standard Horizontal Parameters: Horizontal freq period width rise fall delay duty num points ELECTING EASURE ODES 1. Touch Measure → Measure Setup on the menu bar. 2. Under Measure Mode, select the Std Vertical or Std Horizontal buttons.
Page 112
Getting Started Manual The Detailed selection produces cursors, Measure Gate gate posts, a label identifying the parameter being measured, and a level indicator and hysteresis band for at level parameters (not part of Standard Horizontal or Standard Vertical parameters). Figure 6-1.
Page 113
WavePro 7 Zi and Zi-A Oscilloscopes ARKERS ETUP 1. Touch Measure → Measure Setup on the menu bar. 2. Select a Measure Mode: Std Vertical, Std Horizontal, or My Measure. 3. Touch the Show All button (as follows) to display Help Markers for every parameter being measured on the displayed waveform (C2 in the previous examples).
Getting Started Manual 2. Turn off Help Markers for individual parameters by touching the Px tab for the parameter in question. Then uncheck the AlwaysOn checkbox. When you close this dialog, the Help Markers for this parameter are no longer shown. Histograms Creating and Viewing a Histogram PLEASE NOTE THE FOLLOWING:...
Page 115
WavePro 7 Zi and Zi-A Oscilloscopes 10. On the Buffer side of the Histogram tab, touch inside the #Values data entry field and enter a value. 11. On the Scaling side of the Histogram tab, touch inside the #Bins data entry field and enter a value. 12.
Page 116
Getting Started Manual Click the keypad button (as follows) on the slider bar to enter a value using the pop-up numeric keypad. 10. Touch the Find Center and Width button to center the histogram. Or touch inside the Center, then the Width, data entry fields and enter a value using the slider bar at the bottom of the window.
Page 117
WavePro 7 Zi and Zi-A Oscilloscopes ERSISTENCE ISTOGRAM You can create a histogram of a persistence display also by cutting a horizontal or vertical slice through the waveform. You also decide the width of the slice and its horizontal or vertical placement on the waveform.
Getting Started Manual Math Math Traces and Functions Overview With the instrument’s math tools you can perform mathematical functions on a waveform displayed on any channel, or recalled from any of the four reference memories M1 to M4. You can also set up traces F1 to Fx (the number of math functions that can be performed at the same time depends on the software options loaded on your oscilloscope) to do math on parameter measurements P1 to Px (the number of parameters that can...
Page 119
WavePro 7 Zi and Zi-A Oscilloscopes UNCTION ETUP This setup mode allows you to quickly apply frequently used math functions. 1. Touch Math → Math Setup on the menu bar. 2. If there are math functions already assigned to F1 through Fx (the number of math functions that can be performed at the same time depends on the software options loaded on your oscilloscope, see Specifications for details), touch the checkbox for the function you...
Page 120
Getting Started Manual Mode Description Histogram of the values of a parameter Track of each value of a parameter Trend of the values of a parameter over time Both Track and Trend are tools that can be used to plot measurement data and observe variations with respect to time.
WavePro 7 Zi and Zi-A Oscilloscopes Characteristic Trend Track Monitors an No. Trend points are not Evolution in the evenly spaced in time and Frequency Domain therefore cannot be used for an FFT. Monitors the No. Track resets Evolution of a after every Measurement acquisition.
Getting Started Manual 2. Touch a Math function trace button: F1 through Fx (the number of math functions that can be performed at the same time depends on the software options loaded on your oscilloscope, see Specifications for details), a pop-up menu appears. Select FFT from the menu. 3.
Page 123
WavePro 7 Zi and Zi-A Oscilloscopes Since the data at the end of the record is filled data, it is advisable to select a weighting window other than rectangular to minimize the effect of the fill on the resulting spectrum. 8.
Getting Started Manual Analysis WaveScan NTRODUCTION TO WaveScan enables you to search for unusual events in a single capture, or to scan for an event in many acquisitions over a long period of time. You can select from more than 20 search modes (frequency, rise time, runt, duty cycle, etc.), apply a search condition (slope, level, threshold, hysteresis), and begin scanning in a post-acquisition environment.
Page 125
WavePro 7 Zi and Zi-A Oscilloscopes Signal Views WaveScan provides distinct views of your signal: Source view highlights all occurrences of edges that meet your criteria. Scan Overlay places all captured edges one on top of the other in a separate grid.
Getting Started Manual Parameter Measurements Besides parameter measurements made during acquisition, post- acquisition measurements can also be made. The number of parameters available depends on the options loaded on your instrument. Measurements are made only on the events defined by your filter (search criteria).
Page 127
WavePro 7 Zi and Zi-A Oscilloscopes 6. Touch the newly displayed Trend math function trace label to change settings on the Trend dialog (located on the Trend tab at the lower-right side of the screen). Note: Turning off a trace for which trend data is being collected resets the trend.
Getting Started Manual 6. Touch the newly displayed Track math function trace label if you want to change any settings in the Track dialog: Pass/Fail Testing OMPARING ARAMETERS Each Pass/Fail input (Qx) can compare a different parameter result to a user-defined limit (or statistical range) under a different condition.
Page 129
WavePro 7 Zi and Zi-A Oscilloscopes In Dual Parameter Compare mode, your X-Stream oscilloscope can take parameter results measured on two different waveforms and compare them to each other. If desired, set your test to be true if Any or All waveforms fit the criteria set by the comparison condition.
Page 130
Getting Started Manual You can also decide the actions to occur upon your waveforms' passing or failing, by selecting one or all of the following: stop audible alarm print image of display emit pulse (if available) ...
Page 131
WavePro 7 Zi and Zi-A Oscilloscopes 4. Touch the Summary View checkbox to enable a line of text (as follows) showing a concise status of the last waveform and keeping a running count of how many sweeps have passed. 5. Touch inside the Pass If field, and select a Boolean condition from the choices shown.
Page 132
Getting Started Manual 3. Touch inside the Source1 field and select a source from the pop-up menu. 4. Touch inside the Condition field in the main dialog and select ParamCompare. 5. Touch inside the Compare Values field and select All or Any from the pop-up menu.
Page 133
WavePro 7 Zi and Zi-A Oscilloscopes If you choose either WithinDeltaPct or WithinDeltaAbs from the Condition menu (respectively, previous), you also have the choice of setting the limit by means of the statistical buttons at the bottom of the ParamCompare dialog: Comparing Dual Parameters 1.
Page 134
Getting Started Manual 7. Touch inside the Condition field on the ParamCompare part of the dialog and select a math operator from the pop-up menu. 8. Touch inside the Limit field and enter a value, using the pop-up numeric keypad. This value takes the dimension of the parameter that you are testing.
WavePro 7 Zi and Zi-A Oscilloscopes 5. From the Test portion of the dialog, make a selection in the Test is True when group of buttons (as follows). Note: The previous selection is for All In and means the test is False when a single waveform falls outside the mask.
Page 136
Getting Started Manual You can customize the instrument to your needs by using the power of programs such as Excel™, Mathcad™, and MATLAB™, or by scripting in VBS. Whichever method you use, the results appear on the instrument's display together with the signals that you started with. This ability offers tremendous advantages in solving unique problems for a large range of applications, with comparatively little effort from you.
Documenting Your Work Using LabNotebook Introduction to LabNotebook LeCroy's LabNotebook feature extends the documentation capabilities of your oscilloscope. It allows you to create an annotated notebook entry containing all displayed waveforms, the setup of the DSO, and user- supplied annotation. The notebook entry can then be converted to hardcopy format - pdf, rtf, or html - and printed or e-mailed.
Page 138
Getting Started Manual The Enter Notebook Info pop-up is shown where you can provide a title and comments for the entry. By default, the entry is titled with the current date and time. 2. Touch inside the Title field and enter a title, using the pop-up keyboard.
Page 139
WavePro 7 Zi and Zi-A Oscilloscopes The Circle Tool The circle tool enables you to create a circle around a waveform feature that you want to point out. Once you click off, the circle is drawn and you can drag it anywhere on the screen. The Arrow Tool The arrow tool enables you to draw lines with arrowheads for callouts.
Page 140
Getting Started Manual Colors These are the three default colors that you can select for shapes, lines, and text. To use additional colors, touch More. More (Colors) Button When you touch More, a Custom box opens with the default color yellow displayed.
Page 141
WavePro 7 Zi and Zi-A Oscilloscopes Erase Selected Button If you want to erase a drawing object, touch it to select it, and then touch Erase Selected. Erase All Button Touch Erase All to erase all drawn objects and text. Undo Button Touch Undo to discard the last object drawn.
Getting Started Manual Touch Done when you are finished annotating the notebook entry. The name of the entry is shown on the list box in the LabNotebook dialog. Now, you can now create a hardcopy report of your entry, and email or print it.
WavePro 7 Zi and Zi-A Oscilloscopes AVING ETUP IRECTLY TO A Save your setup(s) directly to a file by touching the File button on the Save To section on the dialog. Now, touch inside the Save files in directory field. Use the pop-up keyboard to provide a path to the destination folder.
Getting Started Manual Finally, touch the Recall Now button. Recalling Default Settings Some of the default settings for your WavePro 700 Zi oscilloscope include the following configurations: Channels 1 and 2 are turned on at 50 mV/div, 0V offset, linear interpolation ...
WavePro 7 Zi and Zi-A Oscilloscopes Saving and Recalling Waveforms Saving Waveforms The Save Waveform dialog is used to save displayed waveforms to either a memory trace or to text or binary file formats. Note: This dialog is for saving waveforms. Save S-parameter files, using the Result Actions Dialog.
Page 146
5. If you are saving to a file, touch the Data Format field and select a format type. Binary - saves the file to LeCroy's binary file format. This format is documented in various Remote Control Manuals for LeCroy Oscilloscopes. Selecting Binary results in the smallest possible file size, and is recommended when recalling waveforms to LeCroy instruments.
Page 147
WavePro 7 Zi and Zi-A Oscilloscopes ASCII -Text output file (.txt extension). MATLAB - Text output file compatible with MATLAB (.dat extension). Excel - Text output file compatible with Excel (.csv extension). MathCad - Text output file compatible with Excel (.prn extension).
Getting Started Manual You can enable Auto Save from the Save Waveform dialog by touching one of the Auto Save buttons. Select Wrap (old files overwritten) or Fill (no files overwritten). CAUTION Selecting Fill can quickly use up all the hard disk space on your instrument.
Page 149
WavePro 7 Zi and Zi-A Oscilloscopes 1. Select either Memory or File by touching either respective button on the Recall From portion of the dialog. Note: When Memory is selected, only Source, Destination, and Show on recall controls are shown on the Recall Waveform dialog. When File is selected, many more controls are available.
Getting Started Manual Disk Utilities Use the Disk Utilities dialog to arrange the file/folder structure on your instrument's hard drive. You can delete files, folders, or create new folders. Note: All Disk Utilities can also be accomplished using the standard Microsoft Windows®...
WavePro 7 Zi and Zi-A Oscilloscopes 2. Touch inside the Current folder data entry field and use the pop-up keyboard to enter the path to the folder that contains the file you want to delete. Touch the Browse button and navigate to the folder. 3.
New software options can be added after purchasing a code and then enabling the option on the oscilloscope. Call LeCroy Customer Support to place an order and receive the code. Contact LeCroy for more Software Option information and refer to the Contact LeCroy for Support (on page 177) topic in this manual for details.
WavePro 7 Zi and Zi-A Oscilloscopes You can log all events or errors only. This log can be invaluable when you are creating and debugging remote control applications. Remote Communication Setup If you are connecting the oscilloscope to a network, first contact your Information Systems administrator.
Getting Started Manual 4. Export the contents of the event log to an ASCII text file by touching the Show Remote Control Log button. The Event Logs pop-up is shown. Touch inside the DestFilename data entry field and enter a file name using the pop-up keyboard.
WavePro 7 Zi and Zi-A Oscilloscopes 3. Under Colors, touch the Use Print Colors checkbox if you want the traces printed on a white background. Note: A white background saves printer toner. (You can change the printer colors in the Utilities → Preference Setup → Color dialog.) 4.
Getting Started Manual Press the P front panel button. Then, touch the Print Setup RINT button on the Print flyout menu. Touch Utilities → Utilities Setup and click the Hardcopy tab. 2. On the Utilities Hardcopy dialog, touch the Printer icon. 3.
WavePro 7 Zi and Zi-A Oscilloscopes You can also choose to print the waveform and grids only, the waveform and grids with the dialog, or the entire screen. You can also e-mail your screen images. Printing Similar to Printing a Screen Image and alternatively to File → Print (and using the additional setting from the Hardcopy dialog) choose from either of the following methods to quickly print from your instrument: ...
Getting Started Manual 1. Touch the Colors field and choose Standard to print the screen as it appears, Print to print the waveforms on a white background, or Black & White to print the waveforms in black & white. Note: Choosing Print colors saves printer toner or ink. 2.
Page 159
WavePro 7 Zi and Zi-A Oscilloscopes Touch Utilities → Utilities Setup and click the Hardcopy tab. On the Hardcopy dialog, touch the File button. 1. Touch inside the File Format data entry field and select a graphic file format from the pop-up menu. 2.
Getting Started Manual E-mail The E-mail selection on the Hardcopy tab gives you the option to e-mail your screen images, using either the MAPI or SMTP protocols. Before you output to e-mail from the Utilities dialog, you first have to set up the e- mail server and recipient address in Preference Setup.
WavePro 7 Zi and Zi-A Oscilloscopes Choose DSO window if you want to print the dialogs with the waveforms and grids. Choose Full Screen if you want to print the entire screen. 5. Touch the Print Now button. Auxiliary Output Signals Although previously mentioned in the Auxiliary Output Signal topic in the Connecting section, the Aux Output tab is on the Utilities dialog, and is briefly mentioned again here for the sake of clarity.
Getting Started Manual Pass/Fail - Allows you to set pulse duration from 1 µs to 131 ms; generates a pulse when pass/fail testing is active and conditions are met. Aux Output Off - This function turns off the auxiliary output signal. Please refer to the previous Auxiliary Output Signal topic in the Connecting section for more information.
WavePro 7 Zi and Zi-A Oscilloscopes Date and Time From the Date and Time tab on the Utilities dialog, you can manually set the time and date or get it from the Internet. If you elect to get the date and time from the Internet, you need to have the oscilloscope connected to the Internet through the LAN connector on the side panel.
From the Options tab on the Utilities dialog, you can add or remove software options. Contact LeCroy for more information about software options by referring to the Contact LeCroy for Support (on page 177) topic in this manual for more details, or visit our Website at www.lecroy.com/options.
WavePro 7 Zi and Zi-A Oscilloscopes 2. Touch the Audible Feedback Enable checkbox so that the oscilloscope beeps with each touch of the screen or front panel control. Enabling Auto-calibration You can choose to have your instrument automatically recalibrate itself whenever there is a significant change in ambient temperature.
Getting Started Manual The advantage of selecting Div is that the waveform remains on the grid as you increase the gain; whereas, if Volts is selected, the waveform could move off the grid. Note: Regardless of whether you select Volts or Div, the Offset shown in the channel setup dialog always indicates volts.
Page 167
WavePro 7 Zi and Zi-A Oscilloscopes 3. Choose an e-mail server protocol: MAPI (Messaging Application Programming Interface) is the Microsoft interface specification that allows different messaging and workgroup applications (including e-mail, voice mail, and fax) to work through a single client, such as the Exchange client. MAPI uses the default Windows e-mail application (usually Outlook Express).
Note: Service dialogs are for LeCroy service personnel use only. A security code is required to gain access. Contact LeCroy service help by referring to the Contact LeCroy for Support (on page 177) topic in this manual for more details.
WavePro 7 Zi and Zi-A Oscilloscopes Touch Screen Calibration Start the calibration procedure by touching the Touch-Screen Calibration button. During calibration you are prompted to touch the center of a small cross in five key touch screen locations. PLEASE NOTE THE FOLLOWING: ...
Page 170
Getting Started Manual Severe cases may require a reloading of the base operating system and oscilloscope application. LeCroy provides you with an Acronis recovery application and a backup image in an extra partition on the instrument’s hard drive. PLEASE NOTE THE FOLLOWING: ...
Page 171
Image Echo Workstation. For more definitive documentation refer to their aforementioned documentation resources. LeCroy provides you with a recovery application and a backup image in an extra partition on the instrument’s hard drive. The recovery process is easy to perform, using the Recovery Wizard. After the recovery procedure is done, you must activate Windows, either by Internet connection to Microsoft’s Web site or by telephone.
Page 172
Getting Started Manual Running the Acronis True Image Echo Workstation Recovery Application 1. Connect a network cable to the LAN port on the rear of the oscilloscope if you intend to activate windows through the Internet. 2. Connect a keyboard and a mouse to the oscilloscope. 3.
Page 173
WavePro 7 Zi and Zi-A Oscilloscopes 13. On the Next Selection page, you will be asked if you want to restore another partition or hard disk drive. Select No, I do not and click Next. 14. The Restoration Options page displays additional options, such as what to do after the restoration is complete and error handling.
Reference WavePro 7 Zi and Zi-ASpecifications Note: Specifications are subject to change without notice. Please refer to the LeCroy website at www.lecroy.com or the online help on your LeCroy oscilloscope for detailed specification information. WP7Zi-GSM-E RevA...
WavePro 7 Zi and Zi-A Oscilloscopes CERTIFICATIONS CE Compliant, UL, and cUL Listed CE Declaration of Conformity The oscilloscope meets requirements of EMC Directive 2004/108/EEC for Electromagnetic Compatibility and Low Voltage Directive 2006/95/EEC for Product Safety. EMC Directive EN 61326-1:2006 ...
Getting Started Manual * In order to conform to Radiated Emissions standards, use properly shielded cables on all I/O terminals. ** Meets Performance Criteria "B" limits during the disturbance; product undergoes a temporary degradation or loss of function of performance which is self recoverable.
You can also find contact information for our offices on the LeCroy Web sites shown for the following regions: Contact Your Local LeCroy Office for Sales and Technical Assistance...
Page 178
Getting Started Manual Contact Your Local LeCroy Office for Sales and Technical Assistance Email (Support): psgsupport@lecroy.com applications.indirect@lecroy.com Web Site: www.lecroy.com Web Site: http://www.lecroy.com/europe Korea - Seoul China LeCroy Korea LeCroy Corporation Beijing Office and Service Center 10th fl.Ildong Bldg. Rm. 2001 - Office; Rm. 2002 - Service Center...
Page 179
WavePro 7 Zi and Zi-A Oscilloscopes Contact Your Local LeCroy Office for Sales and Technical Assistance 1366 9400 Fax (Sales and Support): ++ 886 2 8226 1368 Fax (Sales and Support): ++ 81 4 2402 9586 Email (Sales): sales_twn@lecoln.com.tw Web Site: www.lecroy.com/japan...
Page 180
1. GRANT OF LICENSE. 1.1. License Grant. Subject to the terms and conditions of this EULA and payment of all applicable fees, LeCroy grants to you a nonexclusive, nontransferable license (the “License”) to: (a) operate the Software Product as provided or installed, in object code form, for your own internal business purposes, (i) for use in or with an instrument provided or manufactured by LeCroy (an “Instrument”),...
Page 181
(or other similar designation) the License will not take effect, and you will have no right to use or access the Software Product unless you are properly licensed to use a product identified by LeCroy as being eligible for the upgrade ("Underlying Product"). A Software Product labeled as an "upgrade"...
Page 182
Permitted Objective; (C) the information to be gained thereby has not already been made readily available to you or has not been provided by LeCroy within a reasonable time after a written request by you to LeCroy to provide such information;...
Page 183
Support Services will be considered part of the Software Product and will be subject to the terms and conditions of this EULA. LeCroy may use any technical information you provide to LeCroy during LeCroy’s provision of Support Services, for LeCroy’s business...
Page 184
This EULA grants you no rights to use such Content outside of the proper exercise of the license granted hereunder, and LeCroy will not be responsible or liable therefor. 3.2. Intellectual Property Protection. You may not alter or remove...
Page 185
Documentation) and other proprietary information of LeCroy; and any business, marketing or technical information disclosed by LeCroy, or its representatives, or you in relation to this EULA, and either (i) disclosed in writing and marked as confidential at the time...
Page 186
DFARS §227.7202. The use of the Software Product or Documentation by the Government constitutes acknowledgment of LeCroy’s proprietary rights in the Software Product and Documentation. Manufacturer is LeCroy Corporation, 700 Chestnut Ridge Road, Chestnut Ridge, NY 10977 USA.
Page 187
FITNESS FOR ANY PARTICULAR PURPOSE, AND ALL WARRANTIES IMPLIED FROM ANY COURSE OF DEALING OR USAGE OF TRADE. YOU ACKNOWLEDGE THAT NO WARRANTIES HAVE BEEN MADE TO YOU BY OR ON BEHALF OF LECROY OR OTHERWISE FORM THE BASIS FOR THE BARGAIN BETWEEN THE PARTIES. WP7Zi-GSM-E RevA...
Page 188
SUPPORT SERVICES AGREEMENT WITH LECROY, LECROY’S ENTIRE LIABILITY REGARDING SUPPORT SERVICES WILL BE GOVERNED BY THE TERMS OF THAT AGREEMENT. LECROY SHALL NOT BE LIABLE FOR ANY LOSS OF PROFITS, LOSS OF USE, LOSS OF DATA, INTERRUPTION OF BUSINESS, NOR FOR INDIRECT, SPECIAL,...
Page 189
LeCroy, its agents, or employees, but only by an instrument in writing signed by an authorized officer of LeCroy. No waiver by LeCroy of any breach or default of any provision of this EULA by you will be effective as to...
Software Product or other Confidential Information. ® Windows License Agreement LeCroy's agreement with Microsoft prohibits users from running software on LeCroy X-Stream oscilloscopes that is not relevant to measuring, analyzing, or documenting waveforms. WP7Zi-GSM-E RevA...