Sequence Mode Setup
When setting up Sequence Mode, you define the number of fixed-size
segments acquired in single-shot mode (see the instrument specifications
for the limits).
The oscilloscope uses the sequence timebase setting to determine the
capture duration of each segment. Along with this setting, the oscilloscope
uses the number of segments, maximum segment length, and total
available memory to determine the actual number of samples or
segments, and time or points.