Maintenance

The ultrasound transducers may be disinfected with a suitable disinfectant.
*
Dettol
endoscope disinfectant or a solution of 2% glutaraldehyde are suitable. It
is NOT suitable for steam sterilisation or by disinfectants containing sodium
hypochlorite.
The conductive rubber electrodes and sponge covers may be disinfected using
any suitable disinfectant - a solution of 2% glutaraldehyde may be used. They
are NOT suitable for steam sterilization. Disinfectants containing sodium
hypochlorite should not be used.
The unit may be cleaned by wiping over with a damp cloth. The use of abrasive
materials and cleaning solvents should be avoided.
Regularly inspect all treatment leads, cables and connectors for signs of
damage.
The ultrasonic output power should be checked at least annually.
The mains fuses are mounted on the rear panel to the left of the mains inlet.
The unit must be disconnected from the the mains before any attempt is made
to replace these fuses. The fuse holders are of a bayonet type. A suitable
screwdriver should be used to lightly press in the fuse holder cap and give it a
quarter turn anti-clockwise to release it (reverse for the insertion of replacement
fuses). Information on fuse type and rating is given on the rear panel of the unit
and in the Technical Specification of this manual.
If the mains fuses continue to blow then EMS qualified Service personnel must
be called in.
The Combination 850 has an internal NiMh rechargeable battery. This battery
must only be replaced by authorised service personnel.
There are no user serviceable parts inside the unit and it should not be
opened.
Full servicing instructions are available on request.