Start-up
6.4 System settings
Determine start values using auto function
The following commands can be used to trigger automatic determination of the timing data,
and provides rough approximation of the applied timing.
1. Open a test pattern with a clearly defined border (e.g. SMPTE image).
2. Open the "Live mode" menu: Main menu → Servicelevel 2 → Force Mode → Live mode.
3. Select either "For static pictures" or "For motion pictures" in the "Live mode" menu and
confirm using key 2.
Note: For progressive timings with an image refresh rate greater than 60 Hz, only the
items "For static pictures" or "Reset" can be selected in the "Live mode" menu.
4. Execute the "Force"
62
command using key 2.
SCD 19102 / SMD 19102
Instruction for Use, 05/2013