Eizo SCD 19102 Instructions For Use Manual page 42

19" color and grayscale monitor
Hide thumbs Also See for SCD 19102:
Table of Contents

Advertisement

Start-up
6.3 Description of OSD menu
Main menu
42
Function
Tolerance factor
Signal (only for analog
signal)
ADC calibration
Adjustment / setting
range
1 ... 20
Default: 1
RGB
Monochrome
Default:
RGB (for
SCD19102)
Monochrome (for
)
SMD19102
Execute
Instruction for Use, 05/2013
Description
This tool only consid‐
ers the H and V fre‐
quencies.
An increase in the tab
value results in a larger
tolerance band being
defined. Minor frequen‐
cy fluctuations within
this range will not result
in resynchronization
("Auto in progress").
Switch over signal be‐
tween b/w and color
operation
If a monochrome
signal arrives, it is
green on the color
monitor.
To obtain correct
b/w images, set the
"Signal" parameter
to "Monochrome".
Automatically cali‐
brate A/D converter
for the applied video
level
The video level range
of the system is
checked, and the moni‐
tor set accordingly. This
results in optimum ad‐
justment of the individu‐
al RGB A/D converters
to the video source.
The calibration results
in a larger and more
flexible video level
range (e.g., the full
brightness is also ach‐
ieved at 700 mV if the
video level is limited at
this value).
Note: A specific test
pattern and timing are
prerequisites.
The A/D converters
have already been fac‐
tory-set and need not
be readjusted.
SCD 19102 / SMD 19102

Advertisement

Table of Contents
loading

This manual is also suitable for:

Smd 19102

Table of Contents