Main menu
SCD 19102 / SMD 19102
Instruction for Use, 05/2013
Function
Select gamma curve
Signal setting
Switch loop
6.3 Description of OSD menu
Adjustment / setting
range
1 ... 5
Default:
SCD 19102: LUT 1
1 ... 20
Default: 1
Start-up
Description
Gamma curve selec‐
tion:
information on the se‐
lected curve
The LUTs that can be
selected are displayed
in the table in the OSD
menu.
See Monitor adjustment
– LUT
[}
56]
The following four pa‐
rameters are of rele‐
vance for this tool:
•
H-frequency
•
V-frequency
•
V-total
•
Interlaced/non Inter‐
laced
If one of these parame‐
ters changes, the moni‐
tor treats it as a timing
change and initiates re‐
synchronization via
"Auto in progress". To
prevent this from hap‐
pening as a result of
each and every minor
signal disturbance, the
value representing the
permissible number of
faulty or changed bor‐
ders must be increased
in the case of unstable
signal sources.
Note: The higher the
tab setting, the longer it
will take for a desired
timing change to occur
(delayed by a number
of milliseconds).
41