Start-up
6.4 System settings
3. Select either "For static pictures" or "For motion pictures" in line 6 in the "Live mode"
menu and confirm using key 2.
4. Use the "Horizontal blank pixels" input field to shift the left edge of the SMPTE image
pixel-exact to the left internal border of the gray/blue area.
5. Correct the values in the "Horizontal resolution" input field until the right internal border
of the gray/blue area is pixel-exact at the right edge of the SMPTE image.
6. Use the "Vertical blank pixels" input field to shift the top edge of the SMPTE image pixel-
exact to the top internal border of the gray/blue area.
7. Correct the values in the "Vertical resolution" input field until the bottom internal border
of the gray/blue area is pixel-exact at the bottom edge of the SMPTE image.
8. Execute the "Force"
Note: Gray/blue range not visible
If no gray/blue area is visible on the left and right sides of the image, reduce the "hori‐
zontal resolution" until the gray/blue area becomes visible. If no gray/blue area is visible
on the top and bottom sides of the image, reduce the "Vertical resolution" until the gray/
blue area becomes visible.
Timing successfully set
The unknown timing has been successfully set, you can now exit the OSD menu. The moni‐
tor will recognize the newly set timing and set it as saved whenever connected.
Note
Once all settings in Force Mode have been performed, the zoom factor can be set in the
"Position / Zoom" menu.
Saving of several timings
Save the timings in the OSD menu "Servicelevel 2 → User / Custom settings".
Up to five timings can be saved.
1. Select the storage number via "Custom settings number".
2. Save the timing by selecting "Save custom settings" and execute with key 2.
3. The saved timing is now identified by a "1" in "Save custom settings".
The location in the "Save custom settings" number where the 1 is set depends on the
selected "Custom settings number". For example, if you have selected "1" as the "Cus‐
tom settings number", the 1st position of "Save custom settings" us set to "1".
66
command with key 2.
SCD 19102 / SMD 19102
Instruction for Use, 05/2013