Eizo SCD 19102 Instructions For Use Manual page 49

19" color and grayscale monitor
Hide thumbs Also See for SCD 19102:
Table of Contents

Advertisement

Servicelevel 2
Force Mode
SCD 19102 / SMD 19102
Instruction for Use, 05/2013
Function
Adjustment / setting
range
Serial bus
Enable
Disabled
Default: deactivate
Bus address
Slider from 0 ... 10
Default: 1
H-Scaler clip
Slider from 0 ... 100
Motion image im‐
Active
provement
Inactive
Default:
Interlaced signals: Ac‐
tive
Progressive signals:
Inactive
Blind mode
Execute
Start-up
6.3 Description of OSD menu
Description
To configure a network of
several monitors
This function is used to in‐
terconnect several moni‐
tors so that they can all be
calibrated using a single
computer (connected to the
first monitor).
Note: See Serial interface
[}
25]
Definition of bus address
for monitor
Enables the adjustment
of the aspect ratio
The picture can be expand‐
ed or compressed in hori‐
zontal direction.
Activation of motion im‐
age improvement
Undesirable comb ef‐
fects can be reduced in
the case of interlaced
signals with PAL and
NTSC resolutions con‐
nected to the "VGA"
and "DVI Analog" signal
inputs.
Undesirable staircase
signals can be reduced
in the case of non-inter‐
laced signals (progres‐
sive) of up to 60 Hz
connected to the "VGA"
and "DVI Analog" signal
inputs.
Note: See Motion image
improvement
[}
67]
Force Mode is a tool for
setting unknown timing
that is not displayed via
Exact Modes that have al‐
ready been stored.
Suitable for direct entry of
Force Mode data that have
already been determined.
Note: See Monitor adjust‐
ment - Force Mode
[}
57]
49

Advertisement

Table of Contents
loading

This manual is also suitable for:

Smd 19102

Table of Contents