Things To Do After Operating The Tsq System - Thermo Scientific TSQ Series Hardware Manual

Hide thumbs Also See for TSQ Series:
Table of Contents

Advertisement

Table 3. Typical pressure readings for TSQ Vantage and TSQ Quantum Ultra
Conditions
Collision gas off, ion transfer
tube orifice sealed
Collision gas off, ion transfer
tube orifice open
Collision gas set to 2 mTorr, ion
transfer tube orifice open
Table 4. Typical pressure readings for TSQ Quantum Access
Conditions
Collision gas off, ion transfer
tube orifice sealed
Collision gas off, ion transfer
tube orifice open
Collision gas set to 2 mTorr, ion
transfer tube orifice open
If the pressure is high (above 5 × 10
system within the last 30 to 60 minutes, wait an additional 30 minutes and recheck the pressure.
If the pressure is decreasing with time, check the pressure periodically until it is within the typical
pressure range of the mass spectrometer.
If the pressure remains high, your system might have an air leak. If you suspect an air leak, shut
down the system as described in the topic
Make a visual inspection of the vacuum system and vacuum lines for leaks. Check each fitting and
flange on the system for tightness, and tighten the fittings or flanges that are loose. Do not tighten
fittings indiscriminately. Pay particular attention to fittings that have been changed recently or to
fittings that have been subjected to heating and cooling. Make sure that the cover plates of the
vacuum manifold are properly seated.

Things to Do After Operating the TSQ System

After operating the TSQ system, verify the instrument is ready for a period of inactivity by doing
the following:
Flushing the Sample Transfer Line, Sample Tube, and API Probe
Placing the System in Standby Condition
Cleaning the Ion Sweep Cone and Ion Transfer Tube
Emptying the Solvent Waste Bottle
Thermo Scientific
Convectron gauge reading (foreline,
ion transfer tube- skimmer region)
less than 0.05 Torr
0.9 to 1.2 Torr (110 V at 60 Hz)
1.5 to 2.2 Torr (220 V at 50 Hz)
0.9 to 1.2 Torr (110 V at 60 Hz)
1.5 to 2.2 Torr (220 V at 50 Hz)
Convectron gauge reading (foreline,
ion transfer tube- skimmer region)
less than 0.05 Torr
1.2 to 1.8 Torr (110 V at 60 Hz)
1.5 to 2.2 Torr (220 V at 50 Hz)
1.2 to 1.8 Torr (110 V at 60 Hz)
1.5 to 2.2 Torr (220 V at 50 Hz)
-5
Torr in the analyzer region), and you have restarted the
"Shutting Down the System Completely"
4
Daily Operation
Things to Do After Operating the TSQ System
Ion gauge reading
(analyzer region)
-6
Less than 3 × 10
Torr
(after pumping for at
least one hour)
-6
Less than 5 × 10
Torr
-5
Less than 3 × 10
Torr
Ion gauge reading
(analyzer region)
-6
Less than 3 × 10
Torr
(after pumping for at
least one hour)
-6
Less than 5 × 10
Torr
-5
Less than 3 × 10
Torr
on
page
TSQ Series Hardware Manual
48.
59

Advertisement

Table of Contents
loading

Table of Contents