Maintenance
Maintaining the Ion Source Interface Assembly
Reinstalling the Tube Lens and Skimmer (TSQ Quantum Access and
TSQ Quantum Ultra)
1. Reinstall the tube lens into the ion source interface cage (see
2. Reinstall the skimmer into the ion source interface assembly:
78
TSQ Series Hardware Manual
To reinstall the tube lens and skimmer
a. Orient the tube lens so that the lead pin points toward the socket on the tube lens
connection wire in the contact ring support.
b. Insert the lead pin into the socket and firmly press the tube lens into the contact ring
support.
CAUTION Ensure that you do not scratch or nick the skimmer cone.
a. Orient the skimmer so that the lead pin points toward the socket on the tube lens
connection wire in the contact ring support.
b. Insert the lead pin into the socket and firmly press the skimmer into the contact ring
support.
Figure 46. Ion source interface cage of the TSQ Quantum Access and TSQ Quantum Ultra,
showing the skimmer and tube lens lead pin sockets
Figure
Tube lens lead pin socket
46):
Ion source interface
cage
Contact ring support
Skimmer lead pin
socket
Thermo Scientific