Vacuum Service Switch - Thermo Scientific TSQ Series Hardware Manual

Hide thumbs Also See for TSQ Series:
Table of Contents

Advertisement

Table 2. On/Off status of mass spectrometer components, voltages, and gas flows (Sheet 2 of 2)
Mass spectrometer component
Power supply, electron multiplier,
and conversion dynode
Power supply, 8 kV
PS1 power supply, +24 V
PS2 power supply, +5, ±15,
±24 V dc
PS3 power supply, +36,
-28 V dc
Fan, turbomolecular pump
Fan, above manifold
Fan, center wall
Convectron gauge, foreline
Convectron gauge, collision cell
Ion gauge
Thermo Scientific
Electronics
Standby
Off
service
switch in
Service Mode
position
Off
Off
Off
Off
Off
Off
On
On
On
On
On
Off
On
On
Off
On
On
On
On
On
On
On
On
On
On
On
On
On
On
Off
On
On
Off
3
System Shutdown, Startup, and Reset
Turning Off Selected Mass Spectrometer Components
Vacuum service
MS
switch (TSQ
Main Power
Quantum Access)
circuit breaker
in Service Mode
switch in Off
position
position
Off
Off
Off
Off
Off
Off
Off
Off
Off
Off
Off
Off
Off
Off
Off
Off
On
Off
On
Off
Off
Off