Thermo Scientific TSQ Series Hardware Manual page 89

Hide thumbs Also See for TSQ Series:
Table of Contents

Advertisement

Figure 40. Spray cone, ion transfer tube, O-ring, and ion sweep cone of the ion source interface
5. Soak the ion transfer tube in a dilute solution of nitric acid to remove contaminants.
6. Sonicate the ion transfer tube in distilled water.
7. Clean the ion sweep cone by wiping the inside and outside with methanol and a Kimwipe.
8. Remove, clean with methanol, and inspect the Vespel
seated in the spray cone under the entrance end of the ion transfer tube. Replace it if necessary.
9. Reseat the O-ring in the spray cone.
10. Insert the ion transfer tube into the heater block. Rotate the tube as you insert it. After you
insert it, turn the tube clockwise until it is finger tight.
11. Align the gas inlet on the ion sweep cone with the sweep gas supply port on the ion source
mount. Firmly press the ion sweep cone into the ion source mount. See
Figure
12. Reinstall the Ion MAX or Ion MAX-S source housing on the mass spectrometer as described in
the topic
Thermo Scientific
Ion transfer tube
Spray cone
CAUTION When you reinstall the ion transfer tube:
• Ensure that everything is properly aligned to prevent stripping the threads on the ion
transfer tube.
• Be careful not to bend the ion transfer tube. Rotate the tube as you insert it.
42.
"Installing the Ion MAX or Ion MAX-S Ion Source Housing"
Maintaining the Ion Sweep Cone and Ion Transfer Tube
Vespel O-ring
Ion sweep cone
O-ring (P/N 97055-20442) that is
6
Maintenance
Figure 41
and
on
page
65.
TSQ Series Hardware Manual
71