Cleaning Q00 And L0 - Thermo Scientific TSQ Series Hardware Manual

Hide thumbs Also See for TSQ Series:
Table of Contents

Advertisement

Figure 54. Exploded diagram of the Q00/L0 mount assembly
6. Place all components onto a clean, lint-free surface.
CAUTION Do not disassemble the Q00 RF lens assembly any further.

Cleaning Q00 and L0

To clean the Q00 RF lens and L0 lens
1. Clean the Q00 RF lens and L0 lens with a Kimwipe
2. Dry the parts with a rapid stream of nitrogen gas.
3. Inspect each part for contamination and dust. If necessary, repeat the cleaning procedure.
Thermo Scientific
Lead pins
Guide pin hole
Mounting screws
×
(3
)
Q00 RF lens
assembly
CAUTION Do not use abrasives to clean the Q00 RF lens. Do not soak the Q00 RF lens
and L0 lens in solvent.
Note Wear clean gloves to handle the parts after you clean them.
Guide pin
soaked in methanol.
6
Maintenance
Cleaning the Q00 Ion Optics
Q00/L0 mount
and L0 lens
TSQ Series Hardware Manual
85