Maintenance
Cleaning the Q00 Ion Optics

Reassembling the Ion Optics Cage Assembly

1. Reinstall the Q00 assembly on the Q00/L0 mount:
2. Reinstall the Q00/L0 mount assembly in the ion optics cage:
3. Use needle nose pliers to carefully reconnect the top RF lead wire to the top Q00 lead pin and
4. Reinstall the Q0 mount assembly onto the ion optics cage:
86
TSQ Series Hardware Manual
To reassemble and mount the Q00/L0 assembly
a. Orient the Q00 assembly so that the guide pin on the Q00/L0 mount inserts into the
guide pin hole in the Q00 assembly.
b. Secure the Q00 assembly with the 3 Q00 mounting screws.
a. Orient the Q00/L0 mount assembly so that the lead pins are closest to the lead wires in
the ion optics cage.
b. Press the Q00/L0 assembly into the ion optics cage. See
the bottom RF lead wire to the bottom Q00 lead pin.
a. Orient the Q0 mount assembly so that the flat side is on the same side as the two contacts.
b. Secure the Q0 mount assembly on the ion optics cage with the 3 Q0 mounting screws.
See
Figure
53.
Figure
53.