Shutting Down The System Completely - Thermo Scientific TSQ Series Hardware Manual

Hide thumbs Also See for TSQ Series:
Table of Contents

Advertisement

3
System Shutdown, Startup, and Reset

Shutting Down the System Completely

Shutting Down the System Completely
Shut down the TSQ system completely if not in use for an extended period only, or if it must be
shut down for a maintenance or service procedure. For a short period of time, such as overnight or
over weekends, you can place the system in standby. See
Condition."
1. Turn off the flow of liquid from the LC (or other sample introduction device) to the API
2. From the Tune Master window, choose Control > Off (or click the On/Standby/Off button)
3. Place the Electronics service switch, located on the power panel (see
4. On the TSQ Quantum Access, place the Vacuum service switch, located on the power panel
5. Place the Main Power circuit breaker switch, located on the power panel (see
6. Unplug the power cord for the mass spectrometer.
48
TSQ Series Hardware Manual
To shut down the TSQ system completely
source. To turn off the flow of liquid from the LC to the API source, do the following:
a. Choose Start > Programs > Xcalibur > Quantum Tune to open the Tune Master
window.
b. In the Tune Master window, choose Setup > Inlet Direct Control. The Inlet Direct
Control view appears.
c. Click the LC tab and click the Stop (
to turn off the mass spectrometer.
On
Off
Standby
in the Service Mode position.
(see
Figure 35
on
page
46), in the Service Mode position.
Figure
35) in the Off (O) position. When you place the Main Power circuit breaker switch in
the Off (O) position, the following occurs:
• All power to the mass spectrometer is turned off. (All LEDs on the front panel of the mass
spectrometer are off.)
• A capacitor on the Vent Delay PCB provides power to the vent valve for two to four
minutes (to allow the turbomolecular pump to spin down). After the capacitor discharges,
power to the vent valve solenoid shuts off. When power to the vent valve solenoid shuts
off, the vent valve opens and the vacuum manifold is vented to filtered air.
• After about two minutes, the pressure of the vacuum manifold reaches atmospheric
pressure.
CAUTION Allow heated components to cool before servicing them.
"Placing the System in Standby
) button to stop the LC pump.
Figure 34
and
Figure
35),
Figure 34
and