Table 23
Test Specs for Enclosure Port Immunity to RF Wireless Communications Equipment for
Azure detectors.
Test
Band
Frequency
(MHz)
(MHz)
385
380-390
450
430-470
710
745
704-787
780
810
870
800-960
930
1720
1845
1700-1990
1970
2450
2400-2570
5240
5500
5100-5800
d
5783
e
5785
a. For some services, only the up-link frequencies are included.
b. The carrier shall be modulated using a 50% duty cycle square wave signal.
c. As an alternate to FM modulation, 50% pulse modulation at 18 Hz may be used because while it does not represent actual
modulation, it would be worst case.
d. Test Frequency 5783 is applicable to the Power Supply 33V.
e. Test Frequency 5785 is applicable to the Power Supply 19V.
a
Modulation
Service
Pulse Modulation
TETRA 400
FM Modulation
GMRS 460,
±5 kHz deviation 1
FRS 460
kHz sine
LTE
Pulse Modulation
BAND 13, 17
217 Hz
GSM 800/900;
TETRA 800;
Pulse Modulation
iDEN 820;
CDMA 850;
LTE Band 5
GSM 1800;
CDMA 1900;
GSM 1900;
Pulse Modulation
DECT;
217 Hz
LTE Band
1, 3, 4, 25;
UMTS
Bluetooth,
WLAN, 802.11
Pulse Modulation
b/g/n, RFID
217 Hz
2450, LTE
Band 7
WLAN 802.11
Pulse Modulation
a/n
217 Hz
www.vareximaging.com
Max Power
(W)
b
1.8
18 Hz
c
2
b
0.2
b
2
18 Hz
b
2
b
2
b
0.2
Immunity
Distance
Test Level
(m)
(V/m)
2.5
27
2.5
28
2.5
9
2.5
28
1.25
28
1.25
28
2
9
59