Azure 3131Z - 4131Z Reference Manual
WARNING
Caution
Caution
Caution
Connection of the X-ray Detector to the hospital (clinical) network that
includes other equipment could result in previously unidentified risk
to patients, operators or third parties. The responsible organization
should identify, analyze, evaluate and control these risks. Subsequent
changes to the hospital (clinical) network could introduce new risks
and require additional analysis.
Do not operate the X-ray detector in a location with the following
conditions:
Close to fluid or places where fluid is used
Close to heat sources, such as a heater
High temperature environment
High humidity environment
High condensation environment
Extreme cold environment
Dusty environment
Salty or sulfurous environment
Near a vibrating environment
Environment where there is insufficient air circulation to enable the X-ray
detector and power supply to dissipate heat
Ignoring this caution may result in personal injury or damage to
the product.
Monitor the temperature of the X-ray detector, and allow the X-ray
detector to cool down when the temperature of the X-ray detector gets
hot. Ignoring this caution may result in personal injury or damage to
the product.
The temperature at the back surface of the X-ray detector should not
exceed 35ºC when the unit is installed. This may require air flow over
the back surface of the X-ray detector. Humidity levels should be
between 10-90%.
www.vareximaging.com
Before You Begin
v