Rules for ROI top-left location:
5
Starting location X must be divisible by 4.
Rules for ROI dimensions:
6
The width of the ROI must be divisible by 8.
The width of the ROI must be at least 192 pixels.
The height must be at least 32 pixels.
Due to the impact on the readout time, after the Active Area rectangle has been updated, a new
maximum frame rate will be computed. The current frame rate will only be updated if it is
greater than the new maximum frame rate, at which time, the current frame rate will be set to the
maximum frame rate.