1313DXT-I/1515DXT-I/2520DX-I/3030DX-I X-ray Detectors
13.4
RF Immunity
Table 27
Conducted/Radiated RF
Immunity test
IEC 60601-1-2
test level
3 Vrms
Conducted RF
150 kHz to 80
IEC 61000-4-6
MHz
10 V/m
Radiated RF
80 MHz to 2.7
IEC 61000-4-3
GHz
Compliance
Portable and mobile RF communications
equipment should be used no closer to any part of
the detector than the recommended separation
distance calculated from the equation applicable
3 Vrms
to the frequency of the transmitter.
150 kHz to 80
MHz
Recommended Separation Distance:
d
=
80 MHz to 800 MHz
d
=
800 MHz to 2,5 GHz
d
=
Where P is the maximum output power rating of
the transmitter in watts (W) according to the
transmitter manufacturer and d is the
10 V/m
recommended separation distance in meters (m).
80 MHz to 2.7
Field strengths from fixed RF transmitters, as
GHz
determined by an electromagnetic site survey
should be less than the compliance level in each
frequency range
Interference may occur in the vicinity of
equipment marked with the following symbol:
Some degradation of video & increase in video
noise possible.
www.vareximaging.com
Electromagnetic environment
3.5
------ -
P
V
1
3.5
------ -
P
E
1
7
----- -
P
E
1
b
.
a
57