Varex Imaging Azure 3131Z Manual page 37

X-ray detectors
Table of Contents

Advertisement

Azure 3131Z - 4131Z Reference Manual
Performing a Gain Calibration sequence is used to:
To ensure the X-ray detector is properly calibrated, perform
gain calibration.
It is critical for gain calibration to acquire the flat field image(s) within a range that is
large enough to be higher than the background noise created by the X-ray source and
readout electronics of the detector, but lower than the saturation point of the detector.
Acquire flat field images with a median count of approximately 17000 – 27000.
Using larger numbers of calibration frames to capture the flat field image will result in
more accurate calibration. Accumulating a minimum 32 frames for gain calibration
provides optimal image quality.
Gain calibration should typically be performed:
Every 6 months.
Whenever the central beam of the X-ray source moves relative to the detector.
After replacement of an X-ray tube.
The detector should be at a normal operating temperature before a gain calibration is
performed.
If pulsed X-ray is used, see
on detector X-ray synchronization
Gain calibration should be performed in each operating mode.
Once offset and gain calibration are complete with your selected mode, the
Note
X-ray detector is ready to acquire images.
"Section 20.6, Acquisition Timing" on page 33
www.vareximaging.com
an offset calibration prior to
for information
19

Advertisement

Table of Contents
loading

This manual is also suitable for:

Azure 4131z

Table of Contents