Before You Begin
Installation and Environment of Use
WARNING
WARNING
WARNING
WARNING
WARNING
WARNING
WARNING
WARNING
WARNING
iv
The X-ray detector should be installed, maintained, and used by qualified
professional personnel who are trained and qualified in the installation,
maintenance, and use of X-ray equipment.
The X-ray detector should be mounted onto user-supplied equipment
using the holes provided in the integral flange.
Do not operate the X-ray detector in or around flammable gases, gas mixtures,
liquids, chemicals, or other substances. Ignoring this warning may result in
explosion, fire, or electric shock, which may result in severe personal injury,
death, or substantial product damage.
The enclosure of the X-ray detector must be connected to earth ground
through the X-ray detector grounding lug.
Do not connect the X-ray detector to any component or accessory other
than the manufacturer's specified components and accessories.
Ignoring this warning may result in explosion, fire, or electric shock,
which may result in severe personal injury, death, or substantial
product damage.
Do not modify or alter the X-ray detector, its components, or
accessories. Ignoring this warning may result in explosion, fire, or
electric shock, which may result in severe personal injury, death, or
substantial product damage.
The X-ray detector should not be used as a primary barrier to X-rays.
The user is responsible for ensuring the safety of the operators,
bystanders, and the subjects being X-rayed.
The X-ray detector is not designed to control X-ray dose. The system
integrator is responsible for controlling the X-ray radiation.
The X-ray Detector shall only be used with its approved Varex
Imaging accessories and replacement parts. Product certification and
warranty are rendered void if any alteration to the product is made, or
any instruction, warning, or caution is not followed. It is important
that the X-ray Detector is not directly connected to the hospital
(clinical) network. Connection of the X-ray Detector directly with the
hospital (clinical) network may disturb the IT environment and may
influence the effective data throughput. The imaging workstation and
the WLAN access point must comply with one of the following
standards: IEC 60601-1, IEC 61010-1, IEC 60950-1, or IEC 62368-1.