Troubleshooting; Table 20 Problems And Solutions - Varex Imaging Azure 3131Z Manual

X-ray detectors
Table of Contents

Advertisement

Azure 3131Z - 4131Z Reference Manual
22.0

Troubleshooting

Some possible occurrences and their solutions are shown in
Table 20
Problems and Solutions
Problem
Detector causes Electro-Magnetic
Interference
Acquired image is Non-linear or the
pixel are past the saturation point of
dose
Acquired image is completely dark
Acquired image is noisy
Acquired image shows a white or a
black channel
Reorient or relocate the receiving device.
Increase the separation between the equipment.
Consult the manufacturer or field service technician for help.
Check that the factory supplied cables are used.
Ensure that the tube dose is set to the correct settings.
Acquire another image at the recommended median count as
stated in
"Section 13.2, Gain Calibration" on page
If problem persists; unplug the cables, re-attach the cables,
re-establish the link.
If problem persists, consult the manufacturer or field service
technician for help.
Unplug the cables, re-attach the cables, re-establish the link
and acquire another image.
If problem persists, consult the manufacturer or field service
technician for help.
Ensure that surrounding equipment is not interfering and re-
position the detector further away from equipment.
Remove the cables, re-attach the cables, re-establish the link
and acquire another image.
If problem persists, consult the manufacturer or field service
technician for help.
Ensure that the tube dose is set to the correct settings.
Acquire another image at the recommended median count as
stated in
"Section 13.2, Gain Calibration" on page
If problem persists; unplug the cables, re-attach the cables,
re-establish the link.
If problem persists, consult the manufacturer or field service
technician for help.
www.vareximaging.com
Table
20.
Solution
18.
18.
43