20.12 Alternative Power Supply Requirements
It is possible to power the 3131Z and 4141Z X-ray detectors from a power supply other than the
Varex Imaging provided external power supply.
need to be taken into consideration when selecting or designing a power supply for the X-ray
detector. The connector's pin-out is shown in
FGG.0B.304.CLAD52Z (19V) connector is used on Varex Imaging supplied power supplies and
is suggested as the mating connector for customer supplied power.
Caution
Table 18
Lemo FGG.0B.304.CLAD52Z Pin-out
Connector Pin
1
2
3
4
Figure 17 Lemo FGG.0B.304.CLAD52Z Pin-out Schematic
If an alternate external power supply is used or the power is supplied
in some other way, the EMC and safety test results in this manual are
not applicable. The Original Equipment Manufacturer (OEM)
becomes responsible for performing applicable Electromagnetic
Compatibility (EMC) and safety testing.
www.vareximaging.com
Table 18 on page 39
Figure 17, on page
39.
Signal
Ground
Ground
Power +
Power +
and
Table 19 on page 40
A Lemo