Table 13
Guidance and Manufacturer's Declaration of Electromagnetic Immunity
Guidance and Manufacturer's Declaration of Electromagnetic Immunity
The X-ray detector is intended for use in the electromagnetic environment specified below. The installer,
X-ray system manufacturer, or user of the X-ray detector is responsible for the usage condition of the
detector to be within such environment.
Immunity Test
IEC 60601 Test
Electrostatic
Contact: 6 kV
Discharge (ESD)
Air: 8 kV
IEC 61000-4-2
Electrical fast
0.5 kV (AC)
transients (Burst)
1 kV (DC)
IEC 61000-4-4
Transients-Surges
1 kV /'2 kV
IEC 61000-4-5
Power frequency
3 A/m
magnetic field
IEC 61000-4-8
Voltage dips and
–95%/10 ms
short interruptions
–60%/100 ms
IEC 61000-4-11
–30%/500 ms
>–95%/5000 ms
Electromagnetic Environment –
Compliance
Contact: 6 kV
Floors should be made of wood, concrete,
Air: 8 kV
or ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
0.5 kV (AC)
Mains power quality should be that of a
1 kV (DC)
typical commercial and/or hospital
environment.
1 kV /2 kV
Mains power quality should be that of a
typical commercial and/or hospital
environment.
3 A/m
Power frequency magnetic fields should be
at levels characteristic of a typical location
in a typical commercial and/or hospital
environment.
–95%/10 ms
Mains power quality should be that of a
typical commercial or hospital
–60%/100 ms
environment.
–30%/500 ms
If the user of the X-ray detector requires
>–95%/5000 ms
continued operation during power mains
interruptions, we recommend that the
X-ray detector be powered from an
uninterruptible power supply or battery.
www.vareximaging.com
Guidance
43