Table 15 3131Z Ethernet X-Ray Detector Operational Modes Example - Varex Imaging Azure 3131Z Manual

X-ray detectors
Table of Contents

Advertisement

Azure 3131Z - 4131Z Reference Manual
This feature is not available on the 4131Z Ethernet detector.
Note
Programmable ROI is only available through the API. The API commands
Note
for setting ROI are not available in VSP2Express.
See
Table 15
for an example of the Azure 3131Z X-ray detector mode parameters and
features.
See
Table 16 on page 26
and features.
A complete list of modes can be found in the detector's specifications.
Note
Table 15
3131Z Ethernet X-ray Detector Operational Modes Example
Mode Description
1 × 1
4µGy MLD
Full FOV
2 × 2
16µGy MLD
Full FOV
3 × 3
15µGy MLD
Full FOV
1 × 1
48µGy MLD
Programmable FOV
a. Readout time is for full FOV. Programmable ROI modes readout time is dependent on the programmed matrix size.
b. Max Linear Dose is to be considered a typical value with CsI scintillator and RQA5 beam.
for an example of the Azure 4131Z X-ray detector mode parameters
Readout Time
Pixel Binning
a
[ms]
24.6
12.3
8.2
Variable
Max 24.6
www.vareximaging.com
Max Linear Dose
[µGy/frame]
1 × 1
4
2 × 2
16
3 × 3
15
1 × 1
48
Matrix Size
b
3072 × 3072
1536 × 1536
1024 × 1024
Programmable
25