25.0
Programmable ROI with Viva Pro
In
"Section 18.0, Modes Of Operation and Corrections" on page
possibility to modify the ROI of selected modes (programmable modes).
Please refer to the Varex Smart Panel (VSP2) - Software Interface
Note
Specification for more general information on the programmable ROI.
To change the ROI using Viva Pro, after having installed the detector, connect it to the PC and
start Viva Pro. After establishing a link, select the programmable mode that need to be modify
and then go to Acquisitions Settings. The following window will appear:
Figure 22 Viva Pro Acquisition Settings
Select the Full Size button and a new window will appear.
Figure 23 Viva Pro Acquisition Active Area
50
www.vareximaging.com
Azure 3131Z - 4131Z Reference Manual
24, the 3131Z detector has the