Wear latex or nitrile, powder-free gloves to prevent transfer of oil from hands.
n
Do not drop the standard into the sample cup. Roll the standard into the cup to prevent damage.
n
Always return the standards to the case. Standards are unit specific.
n
S
C
AMPLE
HAMBER
Wear latex or nitrile, powder-free gloves to prevent transfer of oil from hands.
n
Keep the cap on the sample chamber except when inserting or removing the sample cup.
n
When left uncapped, the sample chamber temperature may become unstable and/or
water vapor will adsorb on the inner surface of the chamber. Either of these conditions
can affect analysis results.
Avoid laying the chamber cap on a work surface. Debris may collect on the greased surface of the
n
O-ring which can also affect analysis results.
When multiple analysis modules are connected, never interchange sample chamber caps. Caps
n
are unit specific.
1 - 2
C
OMPONENTS
C
AP
AccuPyc II 1345 Installation Instructions and Checklist
134-42875-02 (Rev A) — Jan 2021