4 Operation | 4.11 Using handpieces
NOTE
If no patient is selected, the images stay in the "Swap Tray" and are not
saved permanently. As soon as a patient is selected, these temporary data in
the "Swap Drying" are deleted. When an active patient is logged off (or a new
patient is logged on) in CONEXIO, a query is shown asking whether the im-
ages shall be deleted or saved. Data deleted at this point cannot be restored
subsequently.

4.11 Using handpieces

NOTE
Consult the separate instructions for installation, use, and servicing for in-
formation regarding the installation, use and servicing of the individual hand-
pieces (e.g., turbine handpiece, camera Satelec Mini LED, PiezoLED, etc.).
Injury or damage due to wear.
Irregular running noise, significant vibration, overheating, imbalance or insuffi-
cient grip.
4 Stop working and contact service support.

4.11.1 Setting spray air and spray water

The adjusting screws for spray air and spray water are situated on the under-
side of the table and correspond to the handpiece that is actually connected.
For KaVo Midwest handpieces, spray air and spray water are pre-set suitable at
the factory such that no re-adjustment is required.
4 Use the short adjusting screw ② on the control valve to adjust the spray air.
ð Counterclockwise rotation (seen from below) increases the flow rate.
4 Use the long adjusting screw ① on the control valve to adjust the spray wa-
ter.
ð Counterclockwise rotation (seen from below) increases the flow rate.
CAUTION
2
1
3
82 / 124