7 Safety check - Test instructions | 7.2 Instructions for the safety check
NOTE
A separate measurement is not usually required for type B applied parts. The
applied parts are connected to the casing (see diagram) and included in the
measurement of the leakage current of the casing, whereby the same per-
missible values are applicable.

7.2.4 Functional tests

The following conditions must be met in all function tests:
▪ The basic function of the treatment centre must be guaranteed.
▪ The treatment centre must be fit for use.
▪ There must be no irregularities, noise or abrasion, etc., present.
The following list is for exemplary purposes and makes no claim of being com-
plete.
▪ Functional test of the safety circuits (see diagram below)
▪ Functioning of the master switch of the device
▪ Functioning of the displays
▪ Functional test of the holder switch of the dentist and assistant element
▪ Functional test of the 3F/MF handpiece – seating of the cannula
▪ Functional test of the operating light
▪ Functional test of the suction hoses
▪ Functional test of the foot control
▪ Function of the chair:
- Travel on all axes
- Test of the limit switches
▪ Functional test ...