2 Safety | 2.1 Explosion hazard

2 Safety

The instructions for use are an integral part of the product and must be read
carefully prior to use and must be accessible at all times.
The device may only be used in accordance with the intended use, any other
type of use is not permitted.
NOTE
All serious events occurring in relation to the product must be reported to the
manufacturer and the competent authority of the member state, in which the
user and/or patient resides.

2.1 Explosion hazard

Electrical sparks in the product can lead to explosion or fire.
4 Do not set up or operate the product in areas where there is an explosion
hazard.
The use of medical gases can lead to explosion or fire.
4 Do not used medical gases in the vicinity of the product.

2.2 Infection hazard

Patients, users or third parties could be infected by contaminated medical
devices.
4 Take suitable personal protective measures.
4 Follow the instructions for use of the components.
4 Before initial startup and after each use, reprocess the product and ac-
cessories appropriately.
4 Reprocess the product and accessories appropriately before disposal.
Given the arrangement of the instruments, injury or infections may occur when
reaching for the tray holder or control panel.
4 Be aware of the arrangement of the instruments when you reach for the
tray holder or operating device.
Suspended instruments (S-table) can cause injury.
4 When you move the dentist element, make sure that nobody gets injured.
It is feasible to use, on standardised interfaces, products from other manufac-
turers, which are not equipped with a protective device preventing back-suction
of treatment water via the instruments into the dental treatment centre. Infec-
tion may be caused by back-suction at the instruments.
4 If products from other manufacturers are used at the standardised inter-
faces, you must make sure that the products are equipped with an appro-
priate protective device.
4 Do not use products with no protective device.
If a water consumer is connected to the third-party connector assembly kit (op-
tional), stagnant water may cause reverse contamination with germs.
4 Before starting, rinse all water drain lines while no handpiece is attached (if
applicable).
4 Before initial start-up and after downtimes (weekends, holidays, vacations,
etc.), rinse or purge the air and water lines with air.
4 Mind the H
O
resistance of the water consumer since the water is doped
2
2
with OXYGENAL 6 (concentration up to 0.02 %).
Germ formation can lead to infection.
4 Carry out an intensive germ reduction.
4 Actuate the tumbler filler several times.