Test Protocol For The Safety Check - KaVo ESTETICA E30 Instructions For Use Manual

Hide thumbs Also See for ESTETICA E30:
Table of Contents

Advertisement

Instructions for use ESTETICA E30
7 Safety check - Test instructions | 7.3 Test protocol for the safety check
(Test of safety and function according to §7 section 3 Medical Device Operation Ordinance)
Operator
Test before startup
Recurrent test
Test after repair
Manufacturer:
Equipment:
Serial number:
ID no.:
Test according to:
IEC 62353 : 2014
Protection class:
Power connection: *
¹
PIE
Applied part, type:
Test:
Visual inspection:
Measurements:
Protective earth resistance
Insulation resistance
Equipment leakage current - Alternative measuring
me
th
od
*
²
A
ppl
ie
d
pa
rt leakage current - Alternative
measuring method
Functional test (according to manufacturer
instructions)
Defect/Comment/Assessment
Overall assessment:
No safety or functional defects detected
No immediate risk, detected defects can be remedied in the short term.
Device must be taken out of commission until defects are remedied!
Device fails to meet requirements - Modification / replacement of components/de-commissioning
recommended.
Date / signature
PIE
Permanently installed equipment
NPS
Non-DETACHABLE POWER SUPPLY CORD
DPS
DETACHABLE POWER SUPPLY CORD
*² The limit of the equipment leakage current corresponds to the value defined in IEC 60601 (DIN EN 60601), taking comment 2 from table 2 into consideration
Date this test protocol was created: 2021-07-28

7.3 Test protocol for the safety check

Test Protocol - Safety Check
I
II
NPS
DPS
B
BF
Limit
0,3
optional: Measurement is not to be
10
mA
5
mA
Testing organisation
Name of the test engineer
Date of testing:
next recurrent test required in
6
12
Measuring device used:
Make:
Type:
Ser. no.:
Calibration date
Measured value
performed
mA
mA
113 / 124
months
18
24
Passes test
no
yes