Instructions for use ESTETICA E30
7 Safety check - Test instructions | 7.3 Test protocol for the safety check
(Test of safety and function according to §7 section 3 Medical Device Operation Ordinance)
Operator
□
Test before startup
□
Recurrent test
□
Test after repair
Manufacturer:
Equipment:
Serial number:
ID no.:
Test according to:
IEC 62353 : 2014
Protection class:
Power connection: *
¹
PIE
Applied part, type:
Test:
Visual inspection:
Measurements:
Protective earth resistance
Insulation resistance
Equipment leakage current - Alternative measuring
me
th
od
*
²
A
ppl
ie
d
pa
rt leakage current - Alternative
measuring method
Functional test (according to manufacturer
instructions)
Defect/Comment/Assessment
Overall assessment:
□
No safety or functional defects detected
□
No immediate risk, detected defects can be remedied in the short term.
□
Device must be taken out of commission until defects are remedied!
Device fails to meet requirements - Modification / replacement of components/de-commissioning
□
recommended.
Date / signature
*¹
PIE
Permanently installed equipment
NPS
Non-DETACHABLE POWER SUPPLY CORD
DPS
DETACHABLE POWER SUPPLY CORD
*² The limit of the equipment leakage current corresponds to the value defined in IEC 60601 (DIN EN 60601), taking comment 2 from table 2 into consideration
Date this test protocol was created: 2021-07-28
7.3 Test protocol for the safety check
Test Protocol - Safety Check
I
II
NPS
DPS
B
BF
Limit
0,3
optional: Measurement is not to be
10
mA
5
mA
Testing organisation
Name of the test engineer
Date of testing:
next recurrent test required in
6
12
Measuring device used:
Make:
Type:
Ser. no.:
Calibration date
Measured value
Ω
Ω
performed
mA
mA
113 / 124
months
18
24
Passes test
no
yes
□
□
□
□
□
□
□
□
□
□
□
□