4 Operation | 4.14 Using the camera
Electrical power.
Electrical shock from incorrectly connecting a non-medical system to the USB
interfaces of the device.
4 Connect any IT device to the medical system in accordance with IEC
60601-1.
4 Use USB devices with no additional power supply (USB-powered) only.
4 Applied parts connected to the USB interface of the dentist element must
comply with the requisite insulation.
4 USB-powered devices failing to meet the requisite insulation for applied
parts must be placed appropriately such that direct contact of the USB
device and the patient is excluded.
4 It is not permissible to touch USB-powered devices failing to meet the re-
quisite insulation for applied parts and the patient at the same time.
The treatment centre may be fitted with up to three USB ports. Camera inter-
faces are situated on the underside of the dentist element (TM-table / T-table /
Cart) or in the dentist element (S-table). Only the cameras approved/enclosed
in the delivery by KaVo may be connected to these interfaces. The USB port in
the back is connected directly to the back-of-the-head PC (in the presence of
the corresponding wiring). USB devices meeting the specifications listed above
can be connected to this interface. To use USB devices that have been connec-
ted, it may be necessary to install a suitable driver software on the back-of-the-
head PC.
Getting the USB ports ready for use
4 To run an USB device, connect the USB port in the terminal box of the
treatment centre to an external back-of-the-head PC. Use one or maximally
two USB extension cables 5 m (mat. no. 1.004.6953) as needed.
4 USB devices connected to the dentist element must meet the USB stand-
ards, USB 1.0, 1.1 or 2.0, and consume max. 500 mA of electrical power.

4.14 Using the camera

Also refer to:
Instructions for use ERGOcam One
Also refer to:
Instructions for use DIAGNOcam 2170 U

4.15 Using the screen

Also refer to:
Instructions for use Screen One / Instructions for use Screen HD

4.16 Connector for third-party equipment

The third-party device connector allows dental devices from third-party manu-
facturers to be used on the KaVo treatment centre.
4 Connect the water ① and air ② supply lines to the couplings.
CAUTION