Instructions for use ESTETICA E30
7 Safety check - Test instructions | 7.2 Instructions for the safety check

7.2 Instructions for the safety check

7.2.1 Preparatory measures to be undertaken on the
device
Electrical power.
Death or bodily injury from electric shock.
4 Before servicing, pull the mains plug out of the socket or completely discon-
nect the device from the power to de-energise it!
4 After conversion, check the electrical safety in accordance with DIN EN
62353 [IEC 62353].
4 Turn off the main switch before any maintenance work.
4 Lift the upholstery at the foot end ① and remove from the chair.
4 Remove two screws ② from the cover of the junction box.
4 Take the cover off proceeding in upward direction.

7.2.2 Visual inspection (inspection by examination)

▪ Check the following items in advance:
Has the configuration of the ME device or the ME system been changed
since the last inspection?
▪ Has the change been documented and approved (test documentation of
safety check)?
WARNING
100 / 124