Instructions for use EXPERTtorque E680 - 1.006.4700EXPERTtorque Mini E677 - 1.006.0100 1 User instructions Serial number Legal Manufacturer CE mark according to Medical Devices Directive EC 93/42 Please note the electronic instructions for use Caution: Consult instructions for use EAC conformity mark (Eurasian Conformity) GOST R certification Transportation and storage conditions (Temperature range)
Instructions for use EXPERTtorque E680 - 1.006.4700EXPERTtorque Mini E677 - 1.006.0100 2 Safety | 2.1 Infection hazard 2 Safety The instructions for use are a component of the product and must be read care- fully prior to use and be accessible at all times. The device may only be used in accordance with the intended use, any other type of use is not permitted.
Repairs, servicing and safety checks may only be performed by trained service personnel. The following persons are authorized to do this: ▪ Service technicians of KaVo branches after the appropriate product training ▪ Service technicians of KaVo authorized dealers after the appropriate product...
Instructions for use EXPERTtorque E680 - 1.006.4700EXPERTtorque Mini E677 - 1.006.0100 3 Description of the product | 3.1 Intended use - Proper use 3 Description of the product ▪ EXPERTtorque LUX E680 LN Mat. no. 1.006.4700 ▪ EXPERTtorque Mini LUX E677 LN Mat. no. 1.006.0100 The EXPERTtorque LUX E680 and the EXPERTtorque Mini LUX E677 are dental handpieces in accordance with 21 CFR §...
Instructions for use EXPERTtorque E680 - 1.006.4700EXPERTtorque Mini E677 - 1.006.0100 3 Description of the product | 3.2 Technical Specifications Proper Use: In accordance with these regulations, this medical device may only be used by a properly trained user and for the application described herein. You need to comply with the following: ▪...
Instructions for use EXPERTtorque E680 - 1.006.4700EXPERTtorque Mini E677 - 1.006.0100 4 Startup and shut-down | 4.1 Checking the water quantity 4 Startup and shut-down WARNING Hazard from non-sterile products. Infection hazard for dentist and patient. ▶ Prior to initial startup and after each use, process the product and accessories.
Instructions for use EXPERTtorque E680 - 1.006.4700EXPERTtorque Mini E677 - 1.006.0100 4 Startup and shut-down | 4.2 Checking the pressures 4.2 Checking the pressures NOTICE Contaminated or moist compressed air at the compressed air connec- tion. Premature wear. ▶ Supply dry, clean and uncontaminated compressed air in accordance with ISO 7494-2 only.
Instructions for use EXPERTtorque E680 - 1.006.4700EXPERTtorque Mini E677 - 1.006.0100 5 Operation | 5.1 Attaching the medical device 5 Operation Note At the beginning of each workday, the water-conducting systems should be rinsed for at least 2 minutes (without transmission handpieces being at- tached) and if there is a risk of contamination from reflux or back suction, the system may also need to be rinsed for 20-30 seconds after each patient.
Instructions for use EXPERTtorque E680 - 1.006.4700EXPERTtorque Mini E677 - 1.006.0100 5 Operation | 5.4 Removing the bur or diamond CAUTION Use of burs with worn or damaged shafts. Risk of injury, tool may fall out during treatment. ▶ Never use burs with damaged or worn shafts. CAUTION Danger of injury from bur.
Page 14
Instructions for use EXPERTtorque E680 - 1.006.4700EXPERTtorque Mini E677 - 1.006.0100 5 Operation | 5.4 Removing the bur or diamond ▶ After the bur has stopped rotating, firmly press the push-button with your thumb and simultaneously pull out the bur. 14 / 26...
Instructions for use EXPERTtorque E680 - 1.006.4700EXPERTtorque Mini E677 - 1.006.0100 6 Troubleshooting | 6.1 Cleaning the spray nozzle 6 Troubleshooting 6.1 Cleaning the spray nozzle WARNING Hazard from non-sterile products. Infection hazard for dentist and patient. ▶ Reprocess and sterilize the medical device properly before the next use. CAUTION Hazard from insufficient amount of spray water.
Instructions for use EXPERTtorque E680 - 1.006.4700EXPERTtorque Mini E677 - 1.006.0100 7 Processing steps in accordance with ISO 17664 | 7.1 Preparations at the site of use 7 Processing steps in accordance with ISO 17664 7.1 Preparations at the site of use WARNING Hazard from contaminated products Contaminated products are associated with an infection risk.
7.2.3 Manual drying ▶ Clean the outside and inside with compressed air until no drops of water are visible. ▶ Immediately after drying, lubricate the KaVo medical device with care agents from the KaVo care system. See also: 2 7.4 Care products and systems - Servicing, Page 18 7.3 Automated processing...
Please comply with the instructions for use of the washer disinfector. ▶ In order to prevent impairment of the KaVo medical device, make sure that the inside and outside of the device is dry after the end of the cycle.
▶ Remove the tool from the medical device. ▶ Cover the medical device with the KaVo Cleanpac bag, and place it on the corresponding care adapter. ▶ Press the spray key once for 1-2 seconds.
Damage to the product. ▶ Remove the product from the steam sterilization immediately after the sterilization cycle. The KaVo medical device has a maximum temperature resistance of up to 138 ℃ (280.4 °F). Sterilization parameters: Select a suitable procedure (depending on the available autoclave) from the fol-...
Instructions for use EXPERTtorque E680 - 1.006.4700EXPERTtorque Mini E677 - 1.006.0100 7 Processing steps in accordance with ISO 17664 | 7.7 Storage ▪ Sterilizer with triple pre-vacuum: - at least 3 minutes at 135°C (275 °F) - Drying time: 16 min. ▪...
NSK spray head 1.005.8436 for KaVo Spray NSK service coupling 1.000.8786 for KaVo QUATTROcare Material summary Mat. No. KaVo Spray USA and Canada 2113 A 0.411.9660 QUATTROcare plus Spray USA and 1.005.4524 Canada 2141 P Chuck care set 1.003.1253 22 / 26...
24 months from the date of the invoice, subject to the following conditions: In case of justified complaints, KaVo will honor its warranty with a free replace- ment or repair. Other claims of any nature whatsoever, in particular with re- spect to compensation, are excluded.