4 Operation | 4.4 Moving the patient chair
Changing the chair's position when the safety shutdown is on.
Personal injury.
Damage to the device.
4 Changing the chair position while a safety shutdown is active, do not move
the chair against the active safety circuit.
Pinching from the patient chair.
The safety shutdown of the patient chair is activated by lifting the respective
component. Depending on the patient's body weight and the leverage, more
force can be exerted on the object to be triggered than is necessary to trigger
the switching function.
4 The treatment personnel must move outside of the chair's swivelling range
whenever the chair moves.
To allow the chair to move freely, it can also be moved when the safety circuit
is on. Use this function for repair purposes only.
4 Press and hold down both the "SP" and "LP" keys on the dentist element.
or
4 Press and hold down the "LP/AP" key on the assistant element.
4 Move the chair using the buttons of the wheel buttons of the chair.

4.4 Moving the patient chair

4.5 Moving the dentist element

Damage from overloading the dentist element.
Damage from overloading the dentist element. Exceeding the maximum weight
of more than 2 kg by adding handpieces, accessories, etc., can cause damage.
4 Do not overload the dentist element!
Risk of injury when the dentist element or assistant element is moved.
The patient or office staff may be injured or bruised.
4 Monitor the patient and office staff when moving the dentist or assistant ele-
ment.