Summary of Contents for KaVo EXPERTmatic E10 C

  • Page 1 Instructions for use EXPERTmatic E10 C – 1.007.5560...
  • Page 2 Manufacturer: Distributed by: KaVo Dental GmbH KaVo Dental GmbH Bismarckring 39 Bismarckring 39 88400 Biberach 88400 Biberach Germany Germany Phone +49 7351 56-0 www.kavo.com Fax +49 7351 56-1488...
  • Page 3: Table Of Contents

    Instructions for use EXPERTmatic E10 C – 1.007.5560 Table of contents Table of contents 1 User instructions....................... 2 Safety ..........................2.1 Infection hazard......................7 2.2 Improper use ....................... 7 2.3 Technical condition......................7 2.4 Accessories and combination with other equipment ............8 2.5 Qualification of personnel....................
  • Page 4: User Instructions

    EXPERTmatic is a registered trademark of KaVo Dental GmbH. All other trademarks are property of their respective owners. KaVo Original Factory Repair In the event of a repair, please ship your product to the KaVo Original Factory Repair using www.kavobox.com. KaVo Technical Service...
  • Page 5 Instructions for use EXPERTmatic E10 C – 1.007.5560 1 User instructions Serial number UDI symbol Manufacturer Note: Please note accompanying documents Follow the electronic instructions for use HIBC Code CE mark for medical devices EAC conformity mark (Eurasian Conformity) Medical device, labelling of medical devices...
  • Page 6 Instructions for use EXPERTmatic E10 C – 1.007.5560 1 User instructions CAUTION In cases which – if not prevented – can lead to minor or moderate injury. CAUTION In cases which – if not prevented – can lead to material damage.
  • Page 7: Safety

    4 Do not operate the medical device at eye level. 2.3 Technical condition A damaged product or damaged or NOT KaVo original components could injure patients, users or third parties. 4 Only operate devices or components if they show no signs of damage on the outside.
  • Page 8: Accessories And Combination With Other Equipment

    Repairs, servicing and safety checks may only be performed by trained service personnel. The following persons are authorised to do this: ▪ Service technicians of KaVo branches after the appropriate product training ▪ Service technicians of KaVo authorised dealers after the appropriate product...
  • Page 9 Define the service interval depending on the frequency of use. As a result of the use of NON-KaVo original spare parts during the repair, parts may become detached and injure patients, users or third parties. This may res- ult in aspiration, swallowing of parts and possibly even a risk of suffocation.
  • Page 10: Description Of The Product

    Instructions for use EXPERTmatic E10 C – 1.007.5560 3 Description of the product | 3.1 Intended use 3 Description of the product EXPERTmatic Handpiece E10 C (Mat. no. 1.007.5560) 3.1 Intended use Indications for use: This medical device is: ▪ Intended for dental treatment only. All other types of use of or modifications to the product are not permitted and can be hazardous.
  • Page 11: Transportation And Storage Conditions

    Instructions for use EXPERTmatic E10 C – 1.007.5560 3 Description of the product | 3.3 Transportation and storage conditions Insert Cutter or diamond grinder according to DIN EN ISO 1797 type 2, for use of a bur stop cutter and diamond grinder...
  • Page 12: Startup And Shut Down

    Dispose of and recycle the packaging appropriately in accordance with current packaging law, employing waste management / recycling companies. Comply with the comprehensive return system. KaVo has had its packaging licensed for this purpose. Please comply with the regional public waste-disposal system.
  • Page 13 Instructions for use EXPERTmatic E10 C – 1.007.5560 4 Startup and shut down | 4.1 Checking the amount of water 13 / 28...
  • Page 14: Operation

    Instructions for use EXPERTmatic E10 C – 1.007.5560 5 Operation | 5.1 Attaching the medical device 5 Operation WARNING Rotating dental bur. Cuts, infection and burn injury. 4 Do not touch the dental bur while it is rotating. 4 Remove the dental bur from the handpiece after treatment to avoid injury and infection during storage.
  • Page 15: Inserting The Dental Bur

    CAUTION Dental bur with damaged, worn or deformed shafts. A damaged product or damaged or NOT KaVo original components can injure patients, users or third parties. Injury hazard, dental bur may fall out during treatment. 4 Never use dental burs with damaged, worn or deformed shafts.
  • Page 16: Removing The Dental Bur

    Instructions for use EXPERTmatic E10 C – 1.007.5560 5 Operation | 5.4 Removing the dental bur CAUTION Dental bur with damaged, worn or deformed shafts. Material damage to the chuck system, dental bur is difficult or impossible to re- move from the chuck system.
  • Page 17: Checking For Malfunctions And Troubleshooting

    Malfunction or complete failure. 4 Do not use Vaseline or other grease or oil. 4 Spray a lint-free cloth with KaVo Spray and apply it to the O-rings on the coupling. 4 Press the O-ring between your fingers to form a loop.
  • Page 18: Cleaning The Spray Nozzle

    Instructions for use EXPERTmatic E10 C – 1.007.5560 6 Checking for malfunctions and troubleshooting | 6.2 Troubleshooting 4 Spray a lint-free cloth with KaVo Spray, apply it to the new O-rings and in- sert them in the recesses. 6.2.2 Cleaning the spray nozzle CAUTION Hazard from insufficient amount of spray water.
  • Page 19: Reprocessing Steps In Accordance With Iso 17664-1 / Iso 17664-2

    Malfunction and material damage. 4 Reprocess in a washer disinfector or by hand only. KaVo recommends the following products based on the compatibility of the ma- terials. The microbiological efficacy must be ensured by the disinfectant manu- facturer and proven by an expert opinion.
  • Page 20: Manual Reprocessing

    Instructions for use EXPERTmatic E10 C – 1.007.5560 7 Reprocessing steps in accordance with ISO 17664-1 / ISO 17664-2 | 7.2 Manual Reprocessing 7.2 Manual Reprocessing WARNING Incomplete disinfection. Patients, users or third parties could be infected by contaminated medical devices.
  • Page 21: Manual Drying

    7.2.3 Manual drying Use KaVo DRYspray for subsequent drying of the air, water and gear unit ducts. 4 Cover the medical device with the KaVo Cleanpac bag, and place it on the corresponding care adapter. 4 Hold the can vertically.
  • Page 22: Pre-Cleaning

    NOTE Please comply with the instructions for use of the washer disinfector. 4 In order not to affect the KaVo medical device, make sure that the product is dry on the inside and outside after completion of the cycle. 4 Remove any residual liquids with KaVoDRYspray.
  • Page 23: Servicing With Kavo Spray

    30 minutes of operation. 4 Remove the dental bur from the medical device. 4 Cover the medical device with the KaVo Cleanpac bag, and place it on the corresponding care adapter. 4 Press the spray key for 1 to 2 seconds.
  • Page 24: Packaging

    Instructions for use EXPERTmatic E10 C – 1.007.5560 7 Reprocessing steps in accordance with ISO 17664-1 / ISO 17664-2 | 7.5 Packaging NOTE End the chuck service mode. Option 1: Place the dental handpieces in the QUATTROcare PLUS 2124 A, close the front flap and start the service procedure.
  • Page 25: Storage

    Instructions for use EXPERTmatic E10 C – 1.007.5560 7 Reprocessing steps in accordance with ISO 17664-1 / ISO 17664-2 | 7.7 Storage 7.7 Storage Reprocessed products must be stored appropriately protected from dust in a dry, dark and cool space with a low germ level.
  • Page 26: Optional Aids And Consumables

    Instructions for use EXPERTmatic E10 C – 1.007.5560 8 Optional aids and consumables 8 Optional aids and consumables Available from specialised dental dealers. Material summary Mat. no. INTRA Instrument stand 3.005.5204 Cleanpac 10 units 0.411.9691 Cellulose pad 100 units 0.411.9862 Nozzle needle 0.410.0921...
  • Page 27: Terms And Conditions Of Warranty

    24 months from the date of the invoice, subject to the following conditions: With regard to justified complaints KaVo grants warranty in the form of a free of charge repair or delivery of a replacement. Other claims of any kind whatso- ever, in particular with respect to compensation, are excluded.