4 Operation | 4.12 Using the KL 703 LED in ENDO mode (optional accessory)
Technical specifications for the KL 703 LED in ENDO
mode
NOTE
The technical specifications apply to the KL 703 LED in ENDO mode.
Speed range
Maximum torque
Operating mode
NOTE
The maximal motor load is 30 seconds operating time / 9 minutes pause (full
load at maximal speed).
In practice, pulse loads lasting seconds or pause times lasting seconds or
minutes are realistic, usually without reaching the maximal possible motor cur-
rent. This corresponds to the common working procedure of a dental profes-
sional.

4.12.2 Calling up ENDO mode

4 Take the endomotor INTRA LUX KL 703 LED from the holder.
4 Press the "Additional motor drives" key.
ð The display switches to the "ENDO" menu.
The device exits from the "Endodontics" treatment mode as soon as the INTRA
LUX KL 703 LED endomotor is returned to the holder. The "Endodontics" treat-
ment mode is activated automatically upon the endomotor being taken out
provided "Endodontics", as the treatment mode, was previously ended by re-
placing the endomotor in the holder.
NOTE
The device does not automatically start when Endo mode was left by pressing
the "Motorised auxiliary drives" button, or if Endo mode has never been activ-
ated since the last time the unit was turned on. The service technician can
disable the automatic start.
NOTE
Before using the endomotor, always check the speed and transfer ratio!
S1
S2
S3
Parameters
Transmission factor
Parameter memory
Torque mode
Direction of motor rotation
200 to 3,200 min
2.5 Ncm
S4
Values
1:1, 3:1
P1,P2,P3,P4,P5,P6
TQ-Ctrl, Autoreverse, AutoRev/Forw.
R, L