Safety Shutdown - KaVo ESTETICA E30 Instructions For Use Manual

Hide thumbs Also See for ESTETICA E30:
Table of Contents

Advertisement

Instructions for use ESTETICA E30
4 Operation | 4.3 Adjusting the dental chair
4 Hold down the foot pedal and foot switch "LP", and simultaneously press
any button for an automatic position ("AP 0", "AP 1", "AP 2" or "SP") on the
dentist or assistant element until you hear a beep.
ð The automatic position is saved to the foot switch.

4.3.5 Safety shutdown

To prevent collisions arising from the movement of the patient chair, safety
shutdown switches are installed to protect the patient and practice personnel
from injury and the treatment centre from damage.
Damage to the assistant element and dental chair.
Despite the safety shutdowns, collisions with the dental chair can take place in
certain positions of the assistant element.
4 Keep the assistant element out of the motion range of the patient chair.
4 Always monitor the chair movement.
Pinching from the patient chair.
The safety shutdown of the patient chair is activated by lifting the respective
component. Depending on the patient's body weight and the leverage, more
force can be exerted on the object to be triggered than is necessary to trigger
the switching function.
4 The treatment personnel must move outside of the chair's swivelling range
whenever the chair moves.
Risk of injury when you move the patient or the patient chair.
The patient or practice personnel can be wedged or crushed.
4 When you move the patient or patient chair, position all moving parts, such
as dentist element, assistant element, operating light, screens, etc. out of
the collision range.
The safety shutdowns can be found at the following places on the treatment
centre:
CAUTION
CAUTION
CAUTION