Short instructions for use 6 Setting the handpieces
6 Setting the handpieces
6.1 Adjusting functions
Settings can be changed using the corresponding keys on the control element or on the
foot control.
The motor speed can be set with the foot control only. The direction of motor rotation,
cooling status and light intensity can be set with the foot control or keys on the control
element.
The current settings are saved and activated the next time the handpiece is taken out.
Setting
Motor speed
Light intensity
Direction of motor rotation
Spray status
Saving the values
▶ Press the "Save" key to save the values. Single values or all values can
be saved simultaneously.
ð Saving is acknowledged by a beep.
speed/intensity
▶ Remove the handpiece (e.g turbine, motor) from the holder.
ð The handpiece is active.
▶ Press the foot pedal.
ð The removed handpiece runs at the set speed or intensity.
▶ Changing speed or intensity with the foot pedal.
ð The left stop corresponds to the minimum speed/intensity.
ð The right stop corresponds to the maximum speed/intensity.
Setting the light intensity
Only the spray status and light intensity can be set for air instruments.
▶ Press the button for "light intensity" (S4) to set the light intensity from 1 to 9 (ring
counter).
Actuation
Foot control
Foot control, control keys
Foot control, control keys
Foot control, control keys
Pane
Display
Display
LED control keys
LED control keys