Instructions for use
INTRAmatic Contra-angle 23 ES - 1.004.5380

Advertisement

Table of Contents
loading

Summary of Contents for KaVo INTRAmatic Contra-angle 23 ES 1.004.5380

  • Page 1 Instructions for use INTRAmatic Contra-angle 23 ES - 1.004.5380...
  • Page 2 Distributed by: Manufacturer: KaVo Dental GmbH Kaltenbach & Voigt GmbH Bismarckring 39 Bismarckring 39 D-88400 Biberach D-88400 Biberach Phone +49 (0) 7351 56-0 www.kavo.com Fax +49 (0) 7351 56-1488...
  • Page 3: Table Of Contents

    Contents 3 / 74 Contents 1 User instructions............................7 2 Safety................................ 9 2.1 Description of safety instructions....................... 9 Warning symbol........................9 2.1.1 Structure..........................9 2.1.2 Description of danger levels....................10 2.1.3 2.2 Safety instructions........................... 12 3 Product description..........................21 3.1 Purpose – Proper use ........................22 3.2 Technical data..........................
  • Page 4 Contents 4 / 74 4.1 Check the water volume........................27 5 Operation..............................30 5.1 Attaching the medical device......................31 5.2 Removing the medical device......................31 5.3 Inserting the milling cutters or diamond grinders................32 5.4 Removing the milling tool or diamond grinder................. 34 6 Troubleshooting............................
  • Page 5 7.3 Disinfection............................48 Manual disinfection - external..................... 49 7.3.1 Manual disinfection - internal....................50 7.3.2 Machine disinfection - external and internal............... 51 7.3.3 7.4 Drying.............................. 52 7.5 Care products and systems - Servicing................... 53 Care with KaVo Spray......................55 7.5.1...
  • Page 6 Contents 6 / 74 Care with KaVo SPRAYrotor....................56 7.5.2 Servicing with KaVo QUATTROcare 2104 / 2104A............57 7.5.3 Servicing with KaVo QUATTROcare PLUS................ 59 7.5.4 Servicing with KaVo QUATTROcare 2104A............... 63 7.5.5 7.6 Packaging............................64 7.7 Sterilisation............................65 7.8 Storage............................67 8 Tools and consumables..........................
  • Page 7: User Instructions

    1 User instructions 7 / 74 1 User instructions Dear User Congratulations on purchasing this KaVo quality product. By following the instructions below you will be able to work smoothly, economically and safely. © Copyright by KaVo Dental GmbH Symbols...
  • Page 8 1 User instructions 8 / 74 Can be steam-sterilised at 134 C -1 C / +4 C (273 F -1.6 +7.4 Thermodisinfectable CE mark (European Community). A product bearing this mark meets the requirements of the applicable EC directive. Action request Target group This document is intended for dentists and their assistants.
  • Page 9: Safety

    2 Safety 9 / 74 2 Safety 2.1 Description of safety instructions 2.1.1 Warning symbol Warning symbol 2.1.2 Structure DANGER The introduction describes the type and source of the haz‐ ard. This section describes potential consequences of non-compliance.
  • Page 10: Description Of Danger Levels

    2 Safety 10 / 74 ▶ The optional step includes necessary measures for hazard prevention. 2.1.3 Description of danger levels The safety instructions listed here, together with the three levels of danger will help avert property damage and injury. CAUTION CAUTION indicates a hazardous situation that can cause damage to property or mild to moderate injuries.
  • Page 11 2 Safety 11 / 74 WARNING WARNING indicates a hazardous situation that can lead to serious or fatal injury. DANGER DANGER indicates a maximal hazard due to a situation that can directly cause death or fatal injury.
  • Page 12: Safety Instructions

    2 Safety 12 / 74 2.2 Safety instructions WARNING Hazard from incorrectly reconditioned products. An infection hazard exists from contaminated products. ▶ Take suitable personal protective measures.
  • Page 13 2 Safety 13 / 74 WARNING Hazards for the care provider and the patient. In the case of damage, irregular running noise, excessive vibration, un- typical warming or when the cutter or grinder cannot be held. ▶ Do not use further and notify Service.
  • Page 14 2 Safety 14 / 74 CAUTION Premature wear and malfunctioning from improper storage during long periods of nonuse. Reduced product life. ▶ The medical device should be cleaned, serviced and stored in a dry lo‐ cation, according to instructions, before long periods of nonuse.
  • Page 15 2 Safety 15 / 74 CAUTION Injury or damage due to wear. Irregular running noise, significant vibration, overheating, imbalance or in‐ sufficient grip. ▶ Stop work and seek service support. CAUTION Risk of burn injury from hot handpiece head or hot hand‐ piece lid.
  • Page 16 2 Safety 16 / 74 CAUTION Risk due to incorrectly stored handpiece. Injury and infection caused by chucked cutter or grinder. ▶ After treatment, place the handpiece properly in the cradle, without the cutter or grinder.
  • Page 17 2 Safety 17 / 74 CAUTION Instructions for the safe use of handpieces with electronic micromotors. Electronic micromotors generate much more energy than conventional pneumatic turbines and motors. Given the higher torque and speed, hand‐ pieces that are poorly serviced, damaged or used improperly can overheat which can seriously burn the patient.
  • Page 18 ▶ After treatment, place the instrument properly in the cradle without the tool. Note For safety reasons, we recommend that the tool holder system be checked annually after the warranty period expires. The following individuals are authorized to repair and service KaVo prod‐ ucts:...
  • Page 19 2 Safety 19 / 74 ▪ Technicians at KaVo branches throughout the world ▪ Technicians specially trained by KaVo The following guidelines must be observed to ensure save use of the elec‐ trically driven handpieces: ▪ The service instructions for handpieces must be precisely following when using KAVO spray or QUATTROcare care systems.
  • Page 20 To ensure proper function, the medical device must be set up according to the reprocessing methods described in the KaVo Instructions for Use, and the care products and care systems described therein must be used. KaVo recommends specifying a service interval at the dental office for a licensed shop to clean, service and check the functioning of the medical device.
  • Page 21: Product Description

    3 Product description 21 / 74 3 Product description INTRAmatic handpiece 23 ES, Mat. no. 1.004.5380...
  • Page 22: Purpose - Proper Use

    3 Product description 22 / 74 3.1 Purpose – Proper use Purpose: This medical device is ▪ intended for dental treatment only. All other types of use or alterations to the product are not permitted and can be hazardous. The medical device is intended for the following applications: Removal of carious material, cavity and crown preparations, removal of fillings, processing of tooth and restoration surfaces.
  • Page 23 3 Product description 23 / 74 ▪ the applicable health and safety regulations ▪ the applicable accident prevention regulations ▪ these Instructions for use According to these regulations, the user is required to: ▪ only use equipment that is operating correctly, ▪...
  • Page 24: Technical Data

    3 Product description 24 / 74 3.2 Technical data Drive speed max. 40,000 rpm Speed transmission 1:3.3 1 red ring Pushbutton chuck 1.6 dia. Milling cutters or grinders ISO 1797-1 type 3 3.3 Transportation and storage conditions CAUTION It is hazardous to start up the medical device after it has been stored strongly refrigerated.
  • Page 25 3 Product description 25 / 74 ▶ Prior to start-up, very cold products must be heated to a temperature of 20 C to 25 C (68 F to 77 Temperature: -20°C to +70°C (-4°F to +158°F) Relative humidity: 5% RH to 95% RH absence of condensation Air pressure: 700 hPa to 1060 hPa (10 psi to 15 psi) Protect from moisture...
  • Page 26: Startup And Shut Down

    4 Startup and shut down 26 / 74 4 Startup and shut down WARNING Hazard from non-sterile products. Infection hazard for care provider and patient. ▶ Before first use and after each use, prepare and sterilise the medical device and accessories accordingly. WARNING Disposal of the product in the appropriate manner.
  • Page 27: Check The Water Volume

    4 Startup and shut down 27 / 74 4.1 Check the water volume CAUTION Overheating of the tooth due to lack of cooling water. Thermal damage to the dental pulp. ▶ Adjust the water amount for the spray cooling to a minimum of 50 /min...
  • Page 28 4 Startup and shut down 28 / 74...
  • Page 29 4 Startup and shut down 29 / 74 CAUTION Hazard from insufficient spray water. Insufficient spray water can cause the medical device to overheat and damage the tooth. ▶ Check the spray water channels and clean the spray nozzles with the nozzle needle Mat.
  • Page 30: Operation

    5 Operation 30 / 74 5 Operation CAUTION Removing and attaching the contra-angle handpiece while the device is rotating. Damage to the drive. ▶ Never attach or remove the contra-angle handpiece while the device is rotating!
  • Page 31: Attaching The Medical Device

    5 Operation 31 / 74 5.1 Attaching the medical device ▶ Place the medical device on the (LUX) motor coupling and lock it into place. ▶ Before each treatment, pull on the medical device to see if it is se‐ curely seated in the motor coupling.
  • Page 32: Inserting The Milling Cutters Or Diamond Grinders

    5 Operation 32 / 74 5.3 Inserting the milling cutters or diamond grinders WARNING Use of unauthorised tools. Injury to the patient or damage to the medical device. ▶ Observe manufacturer instructions and use the tool properly. ▶ Only use tools that do not deviate from the specified data.
  • Page 33 5 Operation 33 / 74 CAUTION Injury from using worn tools. Tools can fall out during treatment and harm the patient. ▶ Never use tools with worn shafts. ▶ Follow the instructions for use of the tool. CAUTION Injury hazard from tools. Infections or cuts.
  • Page 34: Removing The Milling Tool Or Diamond Grinder

    5 Operation 34 / 74 ▶ Press the push button firmly with your thumb and simultaneously in‐ sert the milling or grinding tool until it contacts the stop. ▶ Check that the cutter or grinder is securely attached by pulling on it. 5.4 Removing the milling tool or diamond grinder CAUTION Hazard from rotating tools.
  • Page 35 5 Operation 35 / 74 CAUTION Activation of the push-button collet chuck when the tool is rotating. Injuries. Damage to the chucking system. ▶ Never push the push-button when the tool is rotating! ▶ After the tool has stopped rotating, press the push-button down with your thumb and simultaneously remove the tool.
  • Page 36 5 Operation 36 / 74 DANGER Hazard from defective chucking system. The tool can fall out and cause injury. ▶ Pull on the tool to check if the chucking system is functioning properly and that the tool is firmly clamped. Wear gloves or finger stalls when you check, insert or remove the insert to prevent injury and infection.
  • Page 37: Troubleshooting

    6 Troubleshooting 37 / 74 6 Troubleshooting 6.1 Check for malfunctions CAUTION Missing or damaged O-rings. Malfunctions and premature failure. ▶ Make sure that all O-rings are on the coupling and undamaged.
  • Page 38 6 Troubleshooting 38 / 74 CAUTION Heating of the product. Burns or product damage from overheating. ▶ Do not use the product if it is irregularly heated. ▶ The medical device is too hot while idling: Check the amount of cooling air. ▶...
  • Page 39 6 Troubleshooting 39 / 74 ▶ When the speed drops or is uneven: Caring for the medical device. ▶ An O-ring is missing on the motor coupling: Replace O-ring.
  • Page 40: Troubleshooting

    Hazard from improper care of the O-rings. Malfunctions or complete failure of the product. ▶ Do not use Vaseline or other grease or oil. Note The O-rings on the motor coupling may only be lubricated with a cotton ball wetted with KaVo spray.
  • Page 41: Cleaning The Spray Nozzle

    6 Troubleshooting 41 / 74 ▶ Press the O-ring between your fingers to form a loop. ▶ Shove the O-ring to the front, and remove it. ▶ Insert new O-rings into the grooves. 6.2.2 Cleaning the spray nozzle CAUTION Hazard from insufficient amount of spray water. Insufficient spray water can cause the medical device to overheat and damage the tooth.
  • Page 42: Reprocessing Methods According To Iso 17664

    7 Reprocessing methods according to ISO 17664 42 / 74 7 Reprocessing methods according to ISO 17664 7.1 Preparation at the site of use WARNING Hazard from nonsterile products. There is a risk of infection from contaminated medical devices. ▶ Take suitable personal protective measures. ▶...
  • Page 43: Cleaning

    7 Reprocessing methods according to ISO 17664 43 / 74 ▶ The medical device must be dry when transported for reconditioning. ▶ Do not place it in a solution or similar. 7.2 Cleaning CAUTION Malfunctions from cleaning in the ultrasonic unit. Defects in the product.
  • Page 44: Automated External Cleaning

    10 (e.g. Miele G 7781 / G 7881 – Validation was carried out with Programme "VARIO-TD", cleaning agent "neodisher® mediclean", neutral‐ isation agent "neodisher® Z" and rinsing agent "neodisher® mielclear" and only applies to the material compatibility with KaVo products).
  • Page 45: Manual Cleaning - Internal

    Validated manual internal cleaning (removal of residual protein) can only be done with KaVo CLEANspray and KaVo DRYspray. ▶ Cover the medical device with the KaVo CLEANpac bag, and place it on the corresponding care adapter. Press the spray button three times for 2 seconds each time.
  • Page 46 See also: 2 KaVo CLEANspray / KaVo DRYspray Instructions for Use Note KaVo CLEANspray and KaVo DRYspray for manual interior cleaning are only available in the following countries: Germany, Austria, Switzerland, Italy, Spain, Portugal, France, Luxem‐ bourg, Belgium, Netherlands, United Kingdom, Denmark, Sweden, Fin‐...
  • Page 47: Automated Internal Cleaning

    47 / 74 7.2.4 Automated internal cleaning For alternative interior cleaning, KaVo recommends thermodisinfectors in accordance with EN ISO 15883-1, which are operated with alkaline clean‐ ing agents with a pH value of max. 10 (e. g. Miele G 7781 / G 7881 – Vali‐...
  • Page 48: Disinfection

    7 Reprocessing methods according to ISO 17664 48 / 74 7.3 Disinfection WARNING Hazard due to incomplete disinfection. Principally, KaVo recommends carrying out an final disinfection of the un‐ packaged item if complete disinfection cannot be guaranteed without this measure.
  • Page 49: Manual Disinfection - External

    ▶ Only disinfect in a thermodisinfector or manually. 7.3.1 Manual disinfection - external KaVo recommends the following products based on compatibility of the materials. The microbiological efficacy must be ensured by the disinfec‐ tant manufacturer and proven by an expert opinion.
  • Page 50: Manual Disinfection - Internal

    The efficacy of manual internal disinfection must be demonstrated by the manufacturer of the disinfection agent. With KaVo products, use only disin‐ fection agents that have been released by KaVo with respect to the com‐ patibility of materials (e.g. WL-cid / made by ALPRO).
  • Page 51: Machine Disinfection - External And Internal

    7 Reprocessing methods according to ISO 17664 51 / 74 ▶ Immediately after internal disinfection, lubricate the KaVo medical de‐ vice immediately with care agents from the KaVo care system. 7.3.3 Machine disinfection - external and internal KaVo recommends thermodisinfectors in accordance with EN ISO 15883-1, which are operated with alkaline cleaning agents with a pH value of max.
  • Page 52: Drying

    ▶ In order to prevent negative effects on the medical device, make sure that the interior and the exterior of the medical device are dry after completion of the cycle, and then grease it immediately with servicing agents from the KaVo care system. 7.4 Drying Manual Drying ▶...
  • Page 53: Care Products And Systems - Servicing

    7 Reprocessing methods according to ISO 17664 53 / 74 7.5 Care products and systems - Servicing WARNING Sharp cutters or grinders in the medical device. Risk of injury from sharp and/or pointed cutters or grinders. ▶ Remove cutter or grinder.
  • Page 54 Reduced product life. ▶ Perform proper care regularly! Note KaVo only guarantees that its products will function properly when the care products are used that are listed as accessories since they were tes‐ ted for proper use on our products.
  • Page 55: Care With Kavo Spray

    7 Reprocessing methods according to ISO 17664 55 / 74 7.5.1 Care with KaVo Spray KaVo recommends servicing the product after each time it is used, i.e. after each automatic cleaning and before each sterilisation. ▶ Remove the cutter or grinder.
  • Page 56: Care With Kavo Sprayrotor

    ▶ Carry out the servicing according to the instructions in the section, "Care with KaVo Spray". 7.5.2 Care with KaVo SPRAYrotor KaVo recommends servicing the product as part of the reprocessing after each use, i.e. after each cleaning, disinfection, and before each sterilisa‐ tion.
  • Page 57: Servicing With Kavo Quattrocare 2104 / 2104A

    7 Reprocessing methods according to ISO 17664 57 / 74 ▶ Place the product on the appropriate coupling of the KaVo SPRAYro‐ tor and cover it with a Cleanpac bag. ▶ Servicing the product. See also: 2 Instructions for use KaVo SPRAYrotor 7.5.3 Servicing with KaVo QUATTROcare 2104 / 2104A...
  • Page 58 7 Reprocessing methods according to ISO 17664 58 / 74 KaVo recommends servicing the product as part of the reprocessing after each use, i.e. after each cleaning, disinfection, and before each sterilisa‐ tion. ▶ Remove the cutter or grinder. ▶ Servicing the product.
  • Page 59: Servicing With Kavo Quattrocare Plus

    ▶ Subsequently treat with the specified care products and systems. See also: 2 Servicing with KaVo QUATTROcare 7.5.4 Servicing with KaVo QUATTROcare PLUS KaVo recommends servicing the product after each disinfection, and be‐ fore each sterilisation, in the scope of the reprocessing.
  • Page 60 ▶ Remove the cutter or grinder. ▶ Servicing the product in the QUATTROcare PLUS. See also: 2 Instructions for Use KaVo QUATTROcare PLUS 2124 A Servicing the chucking system KaVo recommends cleaning and servicing the chuck system once a week. See also:...
  • Page 61 7 Reprocessing methods according to ISO 17664 61 / 74 2 Instructions for Use KaVo QUATTROcare PLUS 2124 A Note Handpieces must be taken off the service couplings before the chuck service can be started and performed. ▶ Remove the service coupling chuck from the side hatch of the QUAT‐...
  • Page 62 Option 1: Place the dental handpieces in the QUATTROcare PLUS 2124 A, close the front door and start theservice procedure. Option 2: After three minutes with no service procedure running, the de‐ vice automatically switches back to normal service mode. See also: 2 Servicing with KaVo QUATTROcare PLUS...
  • Page 63: Servicing With Kavo Quattrocare 2104A

    7.5.5 Servicing with KaVo QUATTROcare 2104A Programme-controlled cleaning and servicing device for perfect instrument and turbine care. ▶ Remove the cutter or grinder. ▶ Service the product in QUATTROcare PLUS. See also: 2 Instructions for use KaVo QUATTROcare CLEAN 2140 A...
  • Page 64: Packaging

    7 Reprocessing methods according to ISO 17664 64 / 74 Servicing the chucking system KaVo recommends cleaning and servicing the chuck system once a week using the collet servicing program integrated in the device. See also: 2 Instructions for use KaVo QUATTROcare CLEAN 2140 A 7.6 Packaging...
  • Page 65: Sterilisation

    Sterilisation in a steam steriliser (autoclave) in accordance with EN 13060 / ISO 17665-1 CAUTION Premature wear and malfunctions from improper servicing and care. Reduced product life. ▶ Before each sterilisation cycle, service the medical device with KaVo care products.
  • Page 66 Damage to product. ▶ Immediately remove the product from the steam steriliser after the sterilisation cycle! The KaVo medical device has a maximum temperature resistance up to 138 ℃ (280.4 °F). Select a suitable procedure (depending on the available autoclave) from the following sterilisation processes: ▪...
  • Page 67: Storage

    7 Reprocessing methods according to ISO 17664 67 / 74 ▪ Autoclave using the gravity method: - at least 10 minutes at 134 °C -1 °C / +4 °C (273 °F -1.6 °F / +7.4 °F) ▶ Use according to the manufacturer's Instructions for Use. 7.8 Storage Prepared products must be stored, protected from germs (as far as possi‐...
  • Page 68: Tools And Consumables

    8 Tools and consumables 68 / 74 8 Tools and consumables Available from dental suppliers. Material summary Mat. no. Handpiece stand 2151 0.411.9501 Cellulose pad 100 units 0.411.9862 Cleanpac 10 units 0.411.9691 Nozzle needle 0.410.0931 Sterilisable spray hose (cut section) 0.065.5188 Hose 150 mm (item) 0.593.0252 Material summary...
  • Page 69 8 Tools and consumables 69 / 74 Material summary Mat. no. KaVo CLEANspray 2110 P 1.007.0579 KaVo DRYspray 2117 P 1.007.0580 KaVo Spray 2112 A 0.411.9640 ROTAspray 2 2142 A 0.411.7520 QUATTROcare plus Spray 2140 P 1.005.4525...
  • Page 70: Terms And Conditions Of Warranty

    12 months from the date of the invoice, subject to the following condi‐ tions: In case of justified complaints, KaVo will honour its warranty with a free re‐ placement or repair. Other claims of any nature whatsoever, in particular with respect to compensation, are excluded.
  • Page 71 71 / 74 cal or electrical factors deemed abnormal or impermissible in accordance with KaVo's instructions for use or other manufacturer's instructions. The warranty granted does not usually extend to lamps, light conductors made of glass and glass fibres, glassware, rubber parts, and the colourfastness of plastic parts.