Positioning The Patient Chair Manually - KaVo ESTETICA E30 Instructions For Use Manual

Hide thumbs Also See for ESTETICA E30:
Table of Contents

Advertisement

Instructions for use ESTETICA E30
4 Operation | 4.3 Adjusting the dental chair
4 To swing the headrest, turn the locking dial to the left, move the headrest
into position, and turn the dial to the right to lock it.
4 To remove the headrest cushion, remove the screw ②, pull the cushion ①
up slightly, and remove it to the front.

4.3.3 Positioning the patient chair manually

Danger of injury from overload or dynamic load.
The patient chair might collapse.
4 Do not subject the patient chair to a load exceeding its limit (180 kg).
4 Do not subject the patient chair to dynamic loads.
Motorised motion of the patient chair
The patient or treatment personnel can be clamped or crushed.
4 Monitor the patient and treatment personnel when changing the patient's
position.
Positioning the chair and backrest manually using the
dentist or assistant element
Use the following buttons to adjust the chair height and position of the back-
rest:
Key
CAUTION
CAUTION
Function
The chair moves up.
44 / 124

Advertisement