COMTRAXX® CP305 - Control Panel
2

Intended use

Remote alarm indicator COMTRAXX
®
CP305 is used for visual and audible signalling of operating and alarm
messages from the Bender systems MEDICS
®
, ATICS
®
, EDS and RCMS. In MEDICS
®
monitoring systems, the
CP305 meets the requirements of standard DIN VDE 0100-710 with regard to test functions for IT system
monitoring and alarms from changeover devices. IT system monitoring equipment are tested via the test
button and the devices to be tested.
Important display functions
• Normal operation indicator
• Insulation fault
• Overload
• Overtemperature
• Interruption of the system conductor or PE conductor connection of the ISOMETER
®
• Supply line failure
• Power supply fault conditions and changeover device faults
• Device failure
• Test results
• Measured values
The connection between the CP305 and the transfer switching and monitoring modules is implemented with
bus technology. During normal operation, the CP305 indicates the readiness for operation of the system.
CP305 features 12 digital inputs, allowing messages from other technical equipment to be recorded and
displayed on the CP305, such as from medical gases or battery-supported safety power supply systems (BSV
systems). Two relay contacts enable messages to be forwarded to the higher-level building management
system (BMS).
CP305 devices are used in
• Healthcare facilities
• Industrial and office buildings
• Public buildings
Please heed the limits of the area of application indicated in the technical specifications.
Operator control and adjustment of settings is partly done directly on the device or in a web browser.
Any other use than that described in this manual is regarded as improper.
Intended use includes:
• Equipment-specific settings compliant with local equipment and operating conditions
• Observation of all information in the operating manual
• Compliance with test intervals
CP305_D00425_02_M_XXEN/07.2023