Web user interface: Settings (via ETH socket)
Display
The PIN for logging in to the device consists of 4 digits.
The factory setting is "8070".
• PIN (enabled/disabled)
• PIN (change PIN)
Server
The password for the web user interface is alphanumeric.
The factory setting is "admin".
• Administrator (user name admin)
Status (enabled/disabled)
Password (only *** is displayed)
Username (display only)
Write access (display only)
Read access (display only)
• User (username user)
Status (enabled/disabled)
Password (only *** is displayed)
Username (display only)
Write access (display only)
Read access (display only)
• Not logged in
Write access (display only)
Read access (display only)
12.9
Buzzer
Setting options
• Volume (low, normal, high)
• Mute alarm (off, 1...8 h)
End date (mute alarm): Display of date and time at which the mute is automatically cancelled.
When a mute is set, the LED and header on the device turn blue. A message (e.g. Buzzer off1 h) and the icon of
a crossed-out loudspeaker
During muting, the system does not meet the normative requirements for audible and visual alarms.
12.10 Service inspection
Service interval
During the service inspection, the system does not meet the requirements of the standard for audible and
visual alarms.
Service interval
(deactivated, 3 months ... 4 years)
48
CP305_D00425_02_M_XXEN/07.2023
appear in the device header.