Bender COMTRAXX CP305 Manual page 34

Remote alarm indicator for medical locations and other areas
Hide thumbs Also See for COMTRAXX CP305:
Table of Contents

Advertisement

Display CP305: Menu structure
t(off)
• The switch-off delay indicates how long an alarm needs to no longer be present before the message is also
cancelled.
• Setting range 0...600 s
Relays
Carry out operating principle and function for relays 1...2.
• Test Isometer (relay switches for approx. 30 seconds when a test of the assigned devices (insulation
monitoring devices, LIM, GFCI) is triggered using the "TEST" button).
• Alarm addresses
• Buzzer (the relay switches as soon as the buzzer sounds. In this way, other signalling devices can be installed
to indicate pending messages on the CP305.)
• N/C and N/O operation
Interface
You can enter the settings for the various interfaces here. The current parameter settings are displayed.
Ethernet
– DHCP (on, off); IP address (DHCP)
– Manual: assign an IP address for the local network
The automatically assigned IP address for a 1:1 connection cannot be changed.
BMS
– Address (1...90)
– Interval (1...3 s)
– Acknowledgement via interface (on, off)
NFC
NFC can only be activated for a certain time in order to carry out parameter settings.
– Active (1 h, 12 h, off)
– End (timestamp when NFC is deactivated again)
If the NFC interface is activated, this is indicated on the device with
Individual texts
Individually assigned texts such as device name and digital inputs are displayed but cannot be edited here.
Changes are to be made using the web browser and are automatically applied.
History
Delete the saved history in the device and in the web user interface. This step cannot be undone. If you want to
save the history, use the web user interface to export it before deleting it.
34