Bender COMTRAXX CP9 Series Manual
Hide thumbs Also See for COMTRAXX CP9 Series:

Advertisement

Quick Links

Manual
COMTRAXX® CP9xx –
Control Panel
Remote alarm indicator and operator panel
for medical locations and other areas
Softwareversion: V4.5.x
CP9xx_D00349_08_M_XXEN/02.2022

Advertisement

Table of Contents
loading

Summary of Contents for Bender COMTRAXX CP9 Series

  • Page 1 Manual COMTRAXX® CP9xx – Control Panel Remote alarm indicator and operator panel for medical locations and other areas Softwareversion: V4.5.x CP9xx_D00349_08_M_XXEN/02.2022...
  • Page 2 Bender GmbH & Co. KG © Bender GmbH & Co. KG All rights reserved. PO Box 1161 • 35301 Grünberg • Germany Londorfer Straße 65 • 35305 Grünberg • Germany Reprinting only with permission Tel.: +49 6401 807-0 • Fax: +49 6401 807-259 of the publisher.
  • Page 3: Table Of Contents

    Table of Contents 1. Important information ................... 5 How to use this manual ......................5 Technical support: service and support ................6 1.2.1 First level support ......................6 1.2.2 Repair service ........................6 1.2.3 Field service ........................7 Training courses ........................7 Delivery conditions .........................
  • Page 4 Table of Contents 4.2.1 Continuation of CP915 or CP924 installation ............20 Digital inputs ........................22 Commissioning of the CP9xx device ................23 COMTRAXX® user interface of the CP9xx ..............23 5. PROFINET ....................... 26 Data access via PROFINET ....................26 Device assignment for PROFINET ..................
  • Page 5: Important Information

    COMTRAXX® is a registered trademark of Bender GmbH & Co. KG. Terms used This manual explains Bender-specific terms and functions in detail. Familiarisation with general specialist IT and network terminology is considered a prerequisite. These terms are therefore only explained briefly in this manual.
  • Page 6: Technical Support: Service And Support

    This manual has been compiled with great care. It may nevertheless contain errors and mistakes. Bender can- not accept any liability for injury to persons or damage to property resulting from errors or mistakes in this ma- nual.
  • Page 7: Field Service

    **Mo-Thu 7.00 am - 8.00 pm, Fr 7.00 am - 13.00 pm 1.3 Training courses Bender is happy to provide training regarding the use of test equipment. The dates of training courses and workshops can be found on the Internet at www.bender.de > Know-how > Seminars.
  • Page 8: Warranty And Liability

     ced to the market after 13th August 2005 must be taken back by the manufacturer and disposed of properly. For more information on the disposal of Bender devices, refer to our homepage at www.bender.de > Service & support. CP9xx_D00349_08_M_XXEN/02.2022...
  • Page 9: Safety Instructions

    CP9xx alarm indicator and operator panels display alarms, measured values and states of devices. These include, for example: All Bender devices with BMS bus or BCOM interface  Bender devices (PEM, energy meters,…) with Modbus RTU or Modbus TCP interface  Other devices with Modbus RTU or Modbus TCP interface ...
  • Page 10: Product Description

    The "COMTRAXX® CP9xx - Control Panel" and "BCOM" manuals as PDF files. You can download  the manuals at: http://www.bender.de/en > Service & support > Downloads > Manuals For CP915 and CP924  Connecting cables (video cable, Ethernet keystone coupler, USB cable, RJ45 flat patch cable)
  • Page 11: Possible Applications

    Possible applications Monitoring, operation and display of Medical IT systems  Supply systems for medical gases  HVAC systems  Room lighting  Operating theatre lights  Special power supply systems (UPS)  Other systems from different manufacturers  Configuration, diagnosis, service Each panel can be individually manufactured and tailored to the requirements of the user.
  • Page 12: Applications

    The devices can only be operated on the internal BMS bus. BCOM (Ethernet) for new and future Bender systems, such as ISOMETER® iso685-D.  Modbus RTU (RS-485) when operated as a master for Bender devices PEM… with restricted  functionality (full functionality of PEM…5 only via Modbus TCP).
  • Page 13: Process Image

    Product description 3.8.2 Process image The CP9xx alarm indicator and operator panel combines the information from the different interfac- es and makes it available for operation and visualisation via the web user interface of a PC. It acts as a central user interface. In order to be able to identify them, one individual address is avail- able for each device on this interface.
  • Page 14: Mounting, Connection And Commissioning

    Mounting, connection and commissioning 4. Mounting, connection and commissioning Only qualified personnel are permitted to carry out the work necessary to install, commission and run a device or system. Functional earth The device must be earthed. Without connection of the functional earth, the device function is not guaranteed.
  • Page 15: Dimension Diagram

    4.1.1 Dimension diagram Device dimensions Dimensions (mm) ±1 Type COMTRAXX® CP907 176 (7") CP915 386 (15.6") CP924 610 (24") Installation dimensions enclosure Dimensions (mm) Required Type Enclosure installation depth Glass thickness: 3 mm Flush- mounting CP907 Surface- — mounting Flush- CP 915 mounting Flush-...
  • Page 16: Surface-Mounting Enclosure Cp907

    Mounting, connection and commissioning 4.1.3 Surface-mounting enclosure CP907 For surface mounting, the flush-mounting enclosure is mounted in the optionally available surface- mounting enclosure (B95061915). 90.2 Hinge Dimensions in mm Fig. 4.2: Enclosure for surface mounting CP907 Mounting procedure: 1. Assemble surface-mounting enclosure (fit hinges and bracket). 2.
  • Page 17: Connection Of The Cp9Xx Device

    Mounting, connection and commissioning Removing the CP915/CP924 front plate The front plate is removed from the enclosures of the CP915 or CP924 devices using a suction lifter. For this pur- pose, the suction lifter must be placed on the points marked below one after the other and the front plate must be removed until it clicks into place for the first time.
  • Page 18 Mounting, connection and commissioning Remove the device from the built-in flush-mounting enclosure. d d) Fig. 4.4: Mainboard and connections of the CP9xx devices Connection Connection Plug connector digital inputs Termination of Modbus RTU and BMS bus I²C interface USB ports (for touch sensor) DVI output Plug connector to energy storage board (CP907: not included)
  • Page 19 Mounting, connection and commissioning 3. If the CP9xx is located at the beginning or end of the respective bus (Modbus RTU and BMS), the respective terminating switch of the device (7) must be switched to "ON". Termination Modbus RTU BMS bus R_off R_on | R_off R_on Modbus RTU 4.
  • Page 20: Continuation Of Cp915 Or Cp924 Installation

    Mounting, connection and commissioning 4.2.1 Continuation of CP915 or CP924 installation 1. Connect the front plate to the control board and the power supply unit: Connect a USB port (8) to the board for the touch sensor connection on the front plate. Appro- priate connecting cable included in scope of delivery.
  • Page 21 Mounting, connection and commissioning 4. Attach the front plate to the built-in flush-mounting enclosure. – CP915 – CP924 Hold the display unit from above in front of the flush-mounting enclosure. Place the two mounting hooks behind the upper snap locks of the flush-mounting enclosure (see red markings).
  • Page 22: Digital Inputs

    Mounting, connection and commissioning Digital inputs CP9xx devices feature 12 configurable digital inputs. Settings are made via the COMTRAXX® user interface in a browser. + 24 V + 24 V 9 10 11 12 x, y: 1...12 9 10 11 12 Function The following functions can be assigned to the digital inputs: Digital input without function.
  • Page 23: Commissioning Of The Cp9Xx Device

    Commissioning of the CP9xx device 1. Switch on the supply voltage: After switching on, the device performs a start routine. It is completed when the commission- ing page appears n the display. Fig. 4.5: Commissioning page CP9xx device 2. Enter the desired IP address in the 1 line 3.
  • Page 24 Mounting, connection and commissioning The start screen of the COMTRAXX® user interface appears in the browser window. DeviceName CP900 COMTRAXX® 20.12.2020 13:30 (+01:00) H OM E B US O VER V IE W Device info A L A R M S Comtraxx CP900 V4.x 1207990020-Bxxxxxxxx TO OL S...
  • Page 25 BCOM Multiple assignment of addresses The factory setting for the system name on all Bender BCOM devices is "SYSTEM". If several systems with the same system name are integrated into the same net- work, addresses are assigned twice. This leads to transmission errors. Always en- ter a unique BCOM system name during initial configuration.
  • Page 26: Profinet

     Provision of GSDML file  The GSDML file is also available in the download area of our homepage at https://www.bender.de/en > Service & sup- port > Downloads > Software 5.2 Device assignment for PROFINET To make the required measured values or alarm states available via PROFINET, a device assignment must be generated for the PROFINET image.
  • Page 27: Data Modules

    PROFINET 5.3 Data modules The following data modules can be applied to the available 255 slots in the respective PLC or similar system. The various data modules define which data is to be read via a slot. For each data module, it is also possible to set in the respective PLC or similar system whether a process alarm is to be generated.
  • Page 28: Modbus Tcp Server

    Modbus TCP server 6. Modbus TCP server The Modbus TCP server supports the following function codes: Function code 0x03 (Read Holding Registers)  Function code 0x04 (Read Input Registers)  Function code 0x10 (Preset Multiple Registers)  The Modbus TCP server generates a function-related response to requests and sends it back to the Modbus TCP client.
  • Page 29: Troubleshooting

    (see chapter „3.2 Device features" as well as the manual D00418). Frequently asked questions on the Internet You will find FAQs on many Bender devices at: http://www.bender.de/en > Service & support > Rapid assistance > FAQ Maintenance The device does not contain any parts that must be maintained.
  • Page 30: Technical Data

    Technical data 8. Technical data 8.1 Tabular data ( )* = factory setting Insulation coordination CP907 acc. to IEC 60664-1 Rated voltage.................................50 V Overvoltage category ................................ III Pollution degree ................................2 Rated impulse voltage..............................800 V Insulation coordination CP915 acc. to IEC 60664-1 Rated insulation voltage............................
  • Page 31 Technical data Displays, memory Display/Resolution CP907 ........................7" TFT touch display/800 x 480 CP915 ........................15.6" TFT touch display/1280 x 720 CP924 ..................24" TFT touch display/1280 x 720 or 1920 x 1080 E-mail configuration and device failure monitoring................max. 250 entries Individual texts ................unlimited number of texts with 100 characters each Displayable devices ................................
  • Page 32 Technical data Modbus TCP Interface/protocol ........................Ethernet/Modbus TCP Cable length..............................< 100 m Operating mode.................Client for PEM and "third-party devices" assigned Operating mode........Server for access to process image and for Modbus control commands SNMP Interface/protocol ..........................Ethernet/SNMP Versions................................1, 2c, 3 Supported devices ..................query of all devices (channels) possible Trap support ................................
  • Page 33 Technical data Audio Line IN..................................not used Line OUT..................Output to a STEREO playback device via 3.5 mm jack plug Cable length................................< 3 m Device connections Terminal block (L1; N; PE) (for CP915 and CP924 only) Conductor sizes ............................AWG 20…12 Stripping length............................ 10…11 mm rigid/flexible............................
  • Page 34: Standards, Approvals And Certifications

    Technical data Dimensions CP907 (W x H x D)........................226 x 144 x 78 mm CP915 (W x H x D)........................505 x 350 x 92 mm CP924 (W x H x D)........................654 x 441 x 100 mm Weight CP907 ..............................approx.
  • Page 35: Document Revision History

    Technical data Accessories Device series Description Art. No. CP907 Surface-mounting enclosure B95061915 CP915, CP924 B95061911 CP9xx suction lifter CP9xx replacement plug kit B95061910 The suction lifter is required to remove the display. Document revision history Valid from Document Date software State/Changes version version...
  • Page 36 Technical data CP9xx_D00349_08_M_XXEN/02.2022...
  • Page 37: Index

    Index Address 24 WEEE directive 8 Workshops 7 Claims under the warranty 29 Connection - BMS 18 - PROFIBUS DP 19 Connection, preliminary considerations 14 de-energised 9 Disposal 8 Factory settings communication addresses - Communication addresses 24 GSD file 28 Installation and connection 14 Installation, preliminary considerations 14 Intended use 9...
  • Page 40 Bender GmbH & Co. KG PO Box 1161 • 35301 Grünberg • Germany Londorfer Straße 65 • 35305 Grünberg • Germany Tel.: +49 6401 807-0 • Fax: +49 6401 807-259 E-Mail: info@bender.de • www.bender.de BENDER Photos: Bender Archive...