10.4

Alarm and test

Alarm addresses
Display of the addresses/measuring channels whose alarm messages are assigned to the CP305.
The following are shown:
• BMS address/channel
• Device name
• Path to the alarm value
• Alarm group
• Buzzer on/off
• Buzzer profile
• Buzzer delay
• Relay on/off
The alarm addresses are configured in the web user interface.
Alarm groups
Several measuring channels can be consolidated into one alarm group. The configuration is done in the web
user interface.
Test addresses
Display of the stored test addresses. The configuration is done in the web user interface.
You start the test by pressing the TEST button in the footer of the default display screen. Depending on the
scope of the test, it may take a few minutes. The result is displayed after the test.
The TEST button only appears if at least one test address has been set up.
10.5

Settings

The settings of the CP305 are displayed here. You can adjust some device settings directly on the device if you
log in.
Changed settings are highlighted in yellow. The changes are confirmed (green tick) or discarded (red X) in the
header. You can make several changes at the same time on a single page. A pop-up window prevents you from
leaving the page with unsaved changes.
Digital inputs
Make settings for digital inputs IN1...12. Settings can be made channel by channel.
Mode
• Off
• High-active
• Low-active
t(on)
• The response delay indicates the delay with which the alarm is signalled.
• Setting range 0...600 s
COMTRAXX® CP305 - Control Panel
CP305_D00425_02_M_XXEN/07.2023
33