Web user interface: Settings (via ETH socket)
12.2

Relays 1-2

Menu > Settings > Relay
Carry out operating principle and function for relays 1...2.
Operating principle
Normally open / Normally closed
• Alarm addresses: Relay responds to each alarm address created. This is independent of whether the relay
has been individually parameterised in the alarm addresses.
• Buzzer: The relay switches as soon as the buzzer sounds. In this way, other signalling devices can be installed
to indicate pending messages on the CP305.
• Alarm addresses | Buzzer: Relay switches when either an alarm address or the buzzer is active.
• Test Isometer: (relay switches for approx. 30 seconds when a test of the assigned devices (insulation
monitoring devices, LIM, GFCI) is triggered using the "TEST" button).
12.3

Interface

Menu > Settings > Interface
The parameters required for each connected interface are set here:
Ethernet
Make Ethernet settings. The parameterised IP address is displayed.
If DHCP is switched off, the IP address must be set manually.
BMS
Address (1...90)
Interval (1...3 s)
Acknowledgement via interface (on, off)
NFC
Active (off, 1 h, 12 h)
If NFC is activated, the time of automatic deactivation is shown.
If the NFC interface is activated, this is indicated on the device with
12.4

Individual texts

The individual texts assigned here are displayed on the CP305 but cannot be adjusted there.
Because individual texts are not transmitted via the BMS bus, they are not available on other devices.
46
CP305_D00425_02_M_XXEN/07.2023
Function
Alarm addresses
Buzzer
Alarm addresses | Buzzer
Test Isometer
in the header.