Commissioning - Bender COMTRAXX CP305 Manual

Remote alarm indicator for medical locations and other areas
Hide thumbs Also See for COMTRAXX CP305:
Table of Contents

Advertisement

6

Commissioning

Commissioning assistant
During initial putting into service or after a reset to factory settings, the putting-into-service wizard opens for
the basic setup steps.
The putting-into-service wizard launches in English. The header is blue, and the LED is green.
You need to log in to the device (with the PIN via the icon in the header). Only then are changes possible.
Use the drop down menus to change entries. Changes are confirmed (green) or discarded (red) at the top of
the display.
You can exit the putting-into-service wizard at any time
have already been made are not discarded.
Address settings:
If several parameters can be changed in one box, first touch one of the positions. This is now highlighted in
white and can be set to the correct value with the arrow keys. Repeat these steps until the parameters are
correctly set for the address.
Changes are confirmed (green) or discarded (red) at the top of the display.
The following basic settings are requested and can be set:
Display
• Language
• Alignment
BMS
• BMS address (1 = master)
Ethernet
The IP address that is currently set is displayed.
• DHCP (on/off)
• IP address
• Subnet mask
• Standard gateway
"Complete" will complete the initial setup of the device.
COMTRAXX® CP305 - Control Panel
if you click the icon in the header. Changes that
CP305_D00425_02_M_XXEN/07.2023
25